User Manual

4
6
Description of the Display
This section explains the displayed item on the home display.
ECG1, ECG2, BP, SpO
2
, Impedance RESP, CO
2
Waveform Display
Bed / Room ID No.
Displays the 4-digit Bed ID and 3-digit (000~999) Room ID.
Telemetry Channel
(DS-7141, DS-7101LT)
Displays the telemetry channel ID.
Battery Mark
This mark will be displayed when the monitor is operated with the optional battery. It will be
displayed in 3 levels (Full / Medium / Empty).
Indication of Operating Time
Battery Mark
Charged
Condition
Standard Mode Power Saving Mode
Full
3 hours
2 hours 20 min.
3 hours 30 min.
2 hours
40 min.
Medium
2 hours
10 min. 2 hours 40min.
10 min.
Empty
10 min. 10 min.
Reference
For power saving mode, refer to 8. System Configuration Monitor Setup
CAUTION
The above specification applies when measuring ECG and NIBP
(5-min. interval) using a new battery pack.
The battery pack will degrade with the repeated use, which shortens the
usable time.
Bed / Room ID No.
Telemetry Channel
ECG Lead
ECG Size
BP Scale
SpO
2
Size
RESP Size
Respiratory Sweep
Speed
CO
2
Scale
CO
2
Wave RESP Wave ECG Wave BP Wave SpO
2
Wave
Patient Name Event Key Patient Type Battery Mark Date / Time