User Manual

5
15
Room / Bed ID Setup for LAN Communication
This section describes the procedure to set the Room / Bed ID.
The DS-7100 system incorporates Ethernet LAN unit.
The set Room / Bed ID will be remain stored even when the power is turned off.
Room / Bed ID Setup
To connect to a wired network, it is necessary to set the Room / Bed ID.
1. Press the Menu
Admit / Discharge
Bed ID
keys.
2. Enter the password.
Use the numeric keypad to enter the password
The entered number will be displayed as
****
.
3. Set the Room ID.
Enter the Room ID using the alphanumeric keypad.
The keypad can be selected from ABC or QWERTY
arrangement.
The entered ID will be displayed on the upper left of
the screen.
Next, press the Bed ID
key to display the Bed ID
menu.
4. Set the Bed ID.
Enter the Bed ID using the numeric keypad.
The entered ID will be displayed on the upper left of
the screen.
To connect to the wired network, set the ID in the
range from 001 to 048.
5
Suspend Monitoring