CAUTION
?
The bed ID is factory set to 000. If connected to the wired network
with the bed ID unchanged, monitoring on the central monitor will not be
possible.
?
When connecting to a wired network, make sure that there are no other
bedside monitors with the same ID. If there are more than one
bedside monitor with the same bed ID, the duplicated bedside monitors
can not be monitored on the central monitor.
?
When connecting to a wired network, set the bed ID in the range from
001 to 048.
?
There are following restrictions when connecting the DS-7100 system
to the DS-LAN
network.
?
The DS-7100 system is not corresponded to the AU-5500N 8-channel
recorder. The data for the DS-7100 system can not be recorded on
the AU-5500N.
?
When the measurement unit of BP is kPa, BP waveform, BP numeric
data, NIBP numeric data, NIBP list will not be transmitted. These will
be treated as not measured data, and will not be displayed on the
central monitor. Also, alarm limit setup on the central monitor can not
be performed.
?
When the temperature unit is ?F, the temperature data will not be
transmitted. It will be treated as not measured data, and will not be
displayed on the central monitor. Also, alarm limit setup on the central
monitor can not be performed.
?
Arrhythmia alarm of TACHY, BRADY, SLOW_VT, COUPLET, PAUSE
will not be transmitted.
?
For numeric data displayed as
× × ×
, maximum or minimum value of
measurable range will be transmitted.
?
The numeric data displayed as
will be treated as not measured
data.
?
For QRS classification, the S printed on the built-in recorder will be
printed as N on the HR-500 Recorder.
?
When DS-5800N/NX/NX
MB
is used as a central monitor, the following
function will not be displayed.
?
Recall
?
Graphic Trend, Tabular Trend
?
ST Display