User Manual

11
3
Performance
Display
Device : 8.4 inch TFT Color LCD
Control : Touch Screen Type
Waveform Trace : Stationary Trace
Waveform Speed : ECG / SpO
2
/ BP (6.25mm/s, 12.5mm/s, 25mm/s)
RESP / CO
2
(6.25mm/s, 12.5mm/s, 25mm/ s)
Parameter : ECG, RESP, TEMP, SpO
2
(Arterial Oxygen Saturation), BP1, BP2,
NIBP, CO
2
concentration
Recording
Recording Method : Thermal Array Type
Recording Speed : 25mm/s
Recording Waveform : Max. 3 waveforms
Recording Width : 50mm
Operation
Touch Screen : Eight-Wire Resistive Analog Touch Screen
ECG
Lead Type : Wired 3-electrode, 4-electrode, 5-electrode
Frequency Characteristic : 40Hz / 15Hz
Input Impedance : 5M
?
or above
Max. Input Voltage :
±
10mV
Polarization Voltage :
±
825mV or above
Common Mode
Rejection Ratio : 80 dB or above
HR Measurement Range : Adult 0, 12
300bpm
±
3% or
±
5bpm
Neonate 0, 30
300bpm
±
3% or
±
5bpm
Waveform Size Selection : 1/4, 1/2, 1, 2, 4
Defibrillation Proof : Provided
Respiration
Method : Impedance Method
Frequency Characteristic : 1.5Hz (adult, child) / 2.5Hz (neonate)
Current : 100
μ
A or lower
Measurement Range : 0, 4
150Bpm
±
5Bpm
Temperature
Method : Thermistor Method
Probe : only YSI-400 series
Measurement Range : 0
50
?
C
±
0.2
?
C
No. of Channel : 1 channel
SpO
2
(Arterial Oxygen Saturation)
Method : 2 Wavelength Pulse Wave Method
Measurement Range : 0%
100%
Resolution : 1%
Accuracy : Adult at 70
100%
±
2%
at 0
69% not specified
Neonate at 70
100%
±
3%
at 0
69% not specified
Measurement Range : 20
250bpm
±
3bpm
11
Specification / Performance