User Manual

11
9
Display Configuration
Item Selection Default Backup
No. of waveforms
0
6 waveforms
3 waveforms
No. of numeric
data
0
7 numeric data
4 numeric data
Displayed
waveform
ECG1, ECG2, BP, SpO
2
RESP, CO
2
ECG1, SpO
2
, RESP
Displayed numeric
data
HR, BP, NIBP, SpO
2
RESP, RR, CO
2
,
VPC/ST
HR, NIBP, SpO
2
, RR
Enlarged Display ON, OFF OFF
Display
Configuration
Short Trend ON, OFF, Overlap OFF
/△
NOTE
By selecting ON for backup at discharge, the setup item will be stored
even after discharge.
Selecting OFF will initialize the display configuration to the initial setting of
the selected display mode.
System Configuration Menu
Item Selection Default Backup
Wave Select
ECG1, ECG2, BP, SpO
2
RESP, CO
2
ECG1, BP, RESP
Rec. Duration 24 sec., Continuous 24 sec.
Manual
Recording
Delay Time None, 8sec., 16 sec. 8 sec.
Alarm Record ON, OFF, Center OFF
Wave Select
ECG1, ECG2, BP, SpO
2
RESP, CO
2
, Alarm Factor
ECG1, Alarm Factor
Rec. Duration 12, 24 sec. 12 sec.
Alarm Factor
HR (HR / PR / BPR)
Numeric Data, Arrhythmia
HR (HR / PR / BPR)
Arrhythmia
Alarm
Recording
Arrhythmia Record
ASYSTOLE, VF, VT,
SLOW_VT, RUN,
COUPLET, PAUSE,
BIGEMINY, TRIGEMINY,
FREQUENT, TACHY,
BRADY
ASYSTOLE, VF, VT,
SLOW_VT, RUN,
TACHY, BRADY
Periodic Record ON, OFF, Center OFF
Wave Select
ECG1, ECG2, BP, SpO
2,
RESP, CO
2
ECG1, BP, RESP
Rec. Duration 6, 12, 24 sec. 12 sec.
Periodic
Recording
Periodic Interval 5, 10, 15, 30, 60 min. 60 min.
Paper Feed to Top ON, OFF OFF
Paper Feed to End ON, OFF ON
Rec.
Operation
QRS Classification ON, OFF ON
ECG, BP, SpO
2
25, 12.5, 6.25mm/s 25mm/s
Sweep
Speed
RESP, CO
2
25, 12.5, 6.25mm/s 6.25mm/s
Mode Manual, Auto Manual
Start Time 00:00
23:59 22:00
Complete Time 00:00
23:59 7:00
Display
Time Disp. Only, Slightly
Dark, Dark
Dark
Volume
No change, Quiet, Very
quiet, Silence
Very quiet
Night Mode
Alarm Pole ON, OFF OFF
11
Setup Item