Precautions about Connections to Peripheral Devices
In the interest of safe and sufficient performance of this equipment, the connection of other
manufacturers equipment to the monitor is not authorized, unless the connection is explicitly approved
by Fukuda Denshi. It is the users responsibility to contact Fukuda Denshi to determine the
compatibility and warranty status of any connection made to another manufacturers equipment.
WARNING
For the connector with mark, only the peripheral devices specified by
Fukuda Denshi should be connected with the given procedure. Use of an
unspecified device may cause electric shock to the patient and/or operator due
to excessive leakage current.
CAUTION
All the peripheral device connectors on the DS-7100 system are isolated from
the power supply, but the peripheral devices are not isolated. To prevent
danger of electric shock, always position the peripheral devices away from the
patient.
When connecting peripheral devices to DS-7100 system, it is the users responsibility to verify that the
overall system complies with IEC 60601-1-1, “Collateral Standard: Safety Requirements for Medical
Electrical Systems”.
Precautions about the Fuse
DANGER
If the fuse burns out, contact Fukuda Denshi Service Representative. Do not
continue using it as internal damage to the equipment may be considered.
Accessories and Optional Accessories
WARNING
Use only the cables specified by Fukuda Denshi.
Use of other cables may result in increase in emission or decrease in
immunity.