User's Manual

xv
CAUTION
?
For the following situation, measurements will be terminated.
When the measurement time has exceeded 120 seconds for adult, 90
seconds for child, 60 seconds for neonate.
When the inflation value has exceeded 300mmHg for adult, 200mmHg
for child, 150mmHg for neonate.
?
If used with the incorrect patient type, it will not only cause erroneous
measurement, but the inflating level for the adult may be applied to child
or neonate causing dangerous situation to the patient.
?
The 1-minute interval measurement will always start from 00 second.
Pressing the 1min start
key will start the measurement from the next
00 second.
?
The 1-minute interval measurement will automatically stop after 10
minutes or 20 minutes and returns to the previous interval mode setup.
The selection of 10min
/ 20min can be made on the NIBP
configuration menu.
?
The alarm function will be ineffective for the BP value measured by Quick
SYS regardless of the ON/OFF selection of NIBP alarm.
?
BP Monitoring
?
If the SYS value is abnormally high, or DIA is abnormally low, a
resonance may be the cause. If the resonance can not be eliminated by
adjusting the blood pressure filter, check the BP line and use a thick,
short, and hard catheter.
?
When the main power is turned ON, the BP value will not be displayed
until zero balance is performed. Make sure to perform the zero
balance.
?
Each time the blood pressure transducer or tubing is replaced, the zero
balance procedure is required to ensure accurate measurements.
?
The zero balance procedure is required for the following case.
?
When starting the measurement.
?
When the position of the heart has changed due to body movement.
?
When the position of the transducer has changed.
?
When measuring for a long period of time and there is a possibility of
measurement error due to change in ambient temperature, etc.
?
When the connector is connected / disconnected, or transducer is
replaced.
?
If the mean BP display is set to OFF, the mean BP alarm will not be
generated. Also, the mean BP will not be displayed on the tabular
trend. Be cautious when setting the mean BP display OFF.
?
CO
2
Monitoring
?
Perform calibration after 20 minutes when the main power of the
DS-7100 system is turned ON.
?
Do not disconnect the sampling tube during calibration.
If disconnected, calibration will cease.
?
Conduct CO
2
calibration for the following case.
?
When 6 months has elapsed from the last calibration date.
?
When EtCO
2
measurement is not stable or accuracy is degraded
compared with other measuring device.
?
When the patient monitor was not used for a while, or when EtCO
2
was not measured for a while.
?
CO
2
monitoring will cease when the ventilation connector is covered.
?
Use only the specified sampling tube for CO
2
monitoring.
?
Alarm
?
A faint sound will be generated when setting a minimum volume for the
alarm sound, but be cautious not to miss any alarm. Adjust the volume
to a recognizable level.
?
Alarm messages will be displayed according to the priority. (Level 1
Level 2 Level 3 Level 4)
?
For the same alarm level, the alarm message for the newer alarm will be
displayed.
?
The alarm message for the arrhythmia alarm will continue to be
displayed for 30 seconds after the alarm is resolved.