User's Manual

xvi
CAUTION
?
While the LEAD OFF message is displayed, HR alarm and arrhythmia
alarm will not function. Leaving this condition unresolved may result in
missing a sudden change of the patient. Promptly check the electrodes
when this message is displayed.
?
If the alarm with the higher priority occurs during alarm recording, the
recording in process will be ceased and starts the alarm recording with
the higher priority.
?
Whether to use the SEC alarm function and its threshold selection should
be based on the patients clinical indication portent and medical
evaluation.
?
If the SpO
2
alarm and SEC alarm setup is set to OFF, the SEC alarm
integral value will be set to 0.
?
System Configuration
?
When performing telemetry transmission, the numeric data
corresponding to the waveform should be selected for display.
Otherwise, the displayed waveform or numeric data may not be
transmitted.
?
If the time/date is not correctly set, or changed during monitoring,
erroneous condition may occur to NIBP measurement, periodic
recording, trend and NIBP list data.
?
If the time/date is changed during monitoring, patients age will not be
recalculated.
?
When connected to a wired network, the same time/date with the central
monitor will be set.
?
The alarm ON/OFF setup will remain effective even when the power is
turned OFF. Be cautious not to miss any important alarm by leaving the
alarm silenced.
?
When performing telemetry transmission, the numeric data
corresponding to the waveform should be selected for display.
Otherwise, the displayed waveform or numeric data may not be
transmitted.
?
The setup of channel ID and group ID should be performed only by our
service representative. Users should not perform this procedure as
malfunction to the equipment may occur.
?
The Bed ID is factory set to 000. If connected to the wired network with
the ID unchanged, monitoring on the central monitor will not be possible.
?
When connecting to the wired network, verify that the Bed ID does not
duplicate with other bedside monitors. Otherwise, monitoring on the
central monitor for both bedside monitors will not be possible.
?
To connect to the wired network, set the Bed ID in the range from 001 to
048.
?
There are following restrictions when connecting the DS-7100 system to
the DS-LANnetwork.
?
The DS-7100 system is not corresponded to the AU-5500N 8-
channel recorder. The data for the DS-7100 system can not be
recorded on the AU-5500N.
?
When the measurement unit of BP is kPa, BP waveform, BP numeric
data, NIBP numeric data, NIBP list will not be transmitted. These
will be treated as not measured data, and will not be displayed on the
central monitor. Also, alarm limit setup on the central monitor can
not be performed.
?
When the temperature unit is ?F, the temperature data will not be
transmitted. It will be treated as not measured data, and will not be
displayed on the central monitor. Also, alarm limit setup on the
central monitor can not be performed.
?
Arrhythmia alarm of TACHY, BRADY, SLOW_VT, COUPLET,
PAUSE will not be transmitted.
?
For numeric data displayed as × × × , maximum or minimum value
of measurable range will be transmitted.
?
The numeric data displayed as will be treated as not
measured data.
?
For QRS classification, the S printed on the built-in recorder will be
printed as N on the HR-500 Recorder.