xvii
CAUTION
?
When DS-5800N/NX/NX
MB
is used as a central monitor, recall,
graphic trend, tabular trend, and ST measurement function will not
be displayed.
?
When DS-5700 is used as a central monitor, ST measurement function
will not be displayed
?
A delay will occur for the communication with the central monitor. The
delay is about 1.5 seconds for the waveform, about 1.5~4.0 seconds for
the numeric data, and about 1.5~2.0 seconds for the alarm.
?
When waveform and numeric data display is set to OFF, the alarm
generation and tabular/graphic trend will be also set to OFF.
?
Patient Admit / Discharge
?
If you start monitoring a new patient without performing a discharge
procedure for the previous patient, new data will be added to the
previous data which will result in inaccuracy.
?
The setup for the alarm mode and display mode remains stored even
when the power is turned off or when discharging procedure is
performed.
Before monitoring, make sure the current monitoring mode
is suitable for the patient’s condition.
?
Resuming monitoring will resume the alarm in suspension.
?
Arrhythmia Analysis
?
For proper arrhythmia detection and ECG monitoring, verify proper
electrode placement, lead selection, and ECG waveform size. If
necessary, turn ON the AC filter. Improper electrode placement, lead
selection, and ECG waveform size can cause errors in detection.
?
IC Card
?
Use only the specified IC card.
?
Restart the system after reading the setup data from the IC card.
The setup data will become effective after the system is restarted.
?
Maintenance
?
Always operate the touch screen with fingers or touch screen pen. Do
not touch with a pen-point or other hard-edged instruments.
Malfunction of the touch screen or damage may result.
?
Do not clean the touch panel using strong acid
?
A special coating is applied to the surface of the touch panel. Wipe the
surface with the soft cleaning cloth provided as optional accessory or
with commercially available eyeglass cleaning cloth.
?
Clean the equipment frequently so stains can be removed easily.
?
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
?
Do not allow liquids such as alcohol or cleaning solution enter the
monitor or connectors.
?
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
?
Do not polish the housing with abrasive or chemical cleaner.
?
When sterilizing the entire room using a spray solution, pay close
attention not to have liquids get into the monitor or connectors.
?
Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent
and chemicals (cleanser, thinner, toluene, benzine, benzol, and synthetic
detergent for house and furniture), or sharp-edged tools. The surface
resin coating may be damaged, resulting in discoloration, scratches, and
other problems.
?
Do not open the housing.
?
Replace the components periodically as specified.