User's Manual

xxiii
Compliance to the Electromagnetic Immunity (2)
The DS-7100 system is intended for use in the electromagnetic environment specified below. It
should be assured that the DS-7100 system is used in such an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Radiated RF
IEC61000-4-3
3V/m
26MHz 1.0GHz
3V/m Recommended Separation Distance
d = 1.2
P
d = 1.2
P
80MHz800MHz
d = 2.3
P
800MHz2.5GHz
Where P is the maximum output
power rating of the transmitter in watts
(W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note 1 : At 80Mhz and 800MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which this monitor should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating
this monitor.
b
Over the frequency range 150MHz to 80MHz, field strength should be less than 3V/m.