User's Manual
17
6. Measurement
Turn ON the power and the measurement starts.
■ Starting Screen
When the power is turned ON, the channel number configured on the LX-
7120(G) is displayed at the top of the LCD.
Make sure whether the channel number on the
LCD matches the channel number indicated on
the label of the LX-7120(G) and the channel
number configured on the receiving monitor.
This screen automatically moves onto the next
waveform displa
y
screen.
■ Waveform Display Screen
ECG waveform (1CH when using 3-electrode lead cable, 2CH when using
other lead cable), heart rate, pacemaker marker, respirogram, respiration rate,
remaining battery level, and various messages are displayed.
CAUTION
The LX-7120(G) does not have a diagnostic function. Check the
diagnostic function on the receiving monitor.
The LX-7120(G) does not have an alarm function. Check the alarm
function on the receiving monitor.
The ECG waveform size and sweep speed settings displayed on the
LCD of the LX-7120(G) do not interface with the ones displayed on the
screen of the receiving monitor.
The heart rate and respiration rate displayed on the LCD of the LX-
7120(G) may be different from the ones displayed on the receiving
monitor. Because the algorithm of the ECG and respiration rate is
different.
■ Battery Level Check
Check the battery level on the waveform display screen.
Battery
Symbol
Remaining Battery Level
Full
Getting low but still available
Nearly empty
Replace the battery.
A message that prompts the
battery check appears on the
screen of the receiving
monitor.