User's Manual

49
14. Specification
Specification
CAUTION
For quality improvement, specifications are subject to change without prior
notice.
Standard Specification
Size: 60.6(W) x 60.0(H) x 20.6(D)mm (not including the
protrusion
)
Weight:
A
pproximately 85 grams (with battery)
Transmitting
Waveform:
ECG 1CH or 2CH (selectable from the ECG lead
cable), Respiration waveform
ECG Lead cable Type: 3-electrode, 4-electrode, or 5-electrode (Limb+Chest)
lead cable
A
utomatically detect the type after inserting the lead
cable
Transmitting Status
Data:
Electrode Off, Low Battery, Event Switch, Pacemaker
Detection, Channel ID, 64group Codes
LCD: Built-in
Waterproof: IPX8
Power Supply: DC: one (1) 1.5 V “AA” size (“LR06” size) alkaline
battery
Continuous Operating
Time:
A
pproximately 1.5 days
Continuous operating time is assumed when using new “AA” size (“LR06”
size) Alkaline battery specified by Fukuda Denshi.
ECG
Numbers of Lead
Electrode:
3-electrode, 4-electrode, or 5-electrode
(Limb+Chest)
Numbers of Input
Channel:
1CH(3-electrode) or 2CH
A
ccuracy of Sensitivity: 10mm /1mV ±20%
(Display sensitivity on the receiving monitor)
ECG Input Impedance: 5M and above
Maximum Input Voltage: ±5mV and above
Common Mode Rejection
Ratio:
Less than 10mVp-p (95dB and above)
A
ccuracy of Heart Rate
Measurement:
±10% or ±5bpm, whichever is greater
HR Displa
y
Ran
g
e: 0, 12 to 300bpm
(
1bpm step
)
Frequency
Characteristic:
0.5 to 40Hz (within -3dB)