User's Manual

17
6. SpO
2
Monitoring
When the transmitter is used without the ECG lead cable, it will measure only
SpO
2
.
WARNING
When the SpO
2
probe (sensor) is in a connector-off condition, the SpO
2
alarm will not be generated on the receiving monitor. Make sure that the
SpO
2
probe (sensor)/ relay cable is securely connected. If the SpO
2
waveform/numeric data is not displayed, check the patient’s condition and
pay attention not to miss the connector-off condition.
CAUTION
When using the transmitter with only the SpO
2
sensor cable, ECG and
respiration measurements on the receiving monitor shall be turned off to
prevent an erroneous alarm.
There are varieties of SpO
2
sensors depending on the purpose and intended
use for LX-7230N and LX-7230KM. Select a proper one and attach it to the
patient.
CAUTION
SpO
2
sensor for LX-7230N and SpO
2
sensor for LX-7230KM are not
compatible. SpO
2
sensors for LX-7230KM cannot be used on the LX-
7230N transmitter; vice-versa. Make sure to use the correct sensor for
operation.