User's Manual

6. SpO
2
Monitor (LX-7230KM)
37
CAUTION
The accuracy of the SpO
2
measurement may be influenced by abnormal
hemoglobin, such as carbon monoxide hemoglobin (COHb) and
methemoglobin (MetHb). It may be also affected by cardiogreen or
intravascular dyes.
HbCO METHEMOGLOBIN
SpO
2
1% 5% 10% 1% 5% 10%
50% -0.1% -0.7% -1.5% 0.2% 1.3% 3.2%
70% -0.1% -0.7% -1.5% -0.6% -2.3% -3.2%
90% -0.2% -0.8% -1.6% -1.5% -6.0% -9.6%
100% -0.2% -0.8% -1.7% -1.8% -7.5% -12.2%
Note
The value will be displayed
lower than the actual value.
The value may be
displayed higher than the
actual value when the SpO
2
value is around 50%.
The following case may affect the accuracy of SpO
2
and pulse rate
measurement.
Outside light (direct sunlight, fluorescent, light therapy equipment,
surgical light, infrared heat ramp, etc.)
Hypoperfusion
Excessive body motion (patient’s motion)
Pigment injected to the blood for testing
In case of measurement during receiving CPR treatment
Placement of SpO
2
sensor on limb with a blood-pressure cuff, arterial
catheter
External colorant such as nail polish
Abnormally low or high hemoglobin concentration
Venous pulse
Electrosurgery
Influence of electromagnetic waves from other electronics
High-intensity radio waves from cell phone