Users Manual

4. ECG Monitoring
14
CAUTION
The threshold level for HR detection of this equipment and the receiving
monitor changes with ECG waveform size. Set a proper waveform size
for monitoring.
There are some cases when pacemaker pulse cannot be detected
depending on the pacemaker type, pulse voltage, pulse width,
electrode lead type (unipolar, bipolar), electrode placement, or lead
method which causes the pacemaker pulse amplitude to decrease and
disables pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric
blanket noise or excessive AC frequency noise, these may be
erroneously detected and displayed as a pacemaker pulse. In this
case, check the condition of the electrodes and ECG lead cable to
resolve the cause or turn off the pacemaker detection setting on the
receiving monitor.
Time constant of this equipment is shorter than Fukuda Denshi
monitors (direct ECG connection). Therefore, there is a difference in the
ST measurement value between them. Pay attention to the difference
when monitoring a patient from a transmitter or a monitor.
When an electrode is attached on the same location for a long time,
some patients may develop skin irritation. Check the patient’s skin
condition periodically and change the electrode position as required.
The indication for continuous use of an electrode is about one day.
Replace the electrode if the skin contact gets loosen due to perspiring,
etc.
Make sure to use new disposable electrodes. Otherwise, the waveform
quality may become poor and it may fail to perform correct monitoring
When “Check Electrode” message is displayed on the screen of the
receiving monitor or the display of this equipment, check the condition
of the electrodes and ECG lead cable to resolve the cause.
When removing electrodes from the patient, remove them carefully and
slowly. Do not apply excessive force to remove them. Otherwise, it may
damage the skin.
A correct measurement may not be performed depending on the
attached position of the electrodes. Attach the electrodes on the patient
referring to “Connecting the ECG Lead Cable and Electrodes” and
make sure that the correct waveform is measured on the display.