Instruction Manual
- 7 - 
Tel: (949) 465-0900 
Fax: (949) 465-0905 
Toll Free: (800) 23 FUTEK 
Ex. If you are using your CSG110 with a tension and compression load cell and you 
have tension setup as the positive direction, but now you would like to have 
compression as the positive direction, all you have to do is flip the DIP switch from the 
default polarity position to the reverse polarity position. 
UP reverse
DOWN straight
SW2: Polarity
Gain 
There are eight gain settings available on the CSG110: from 0.5 mV/V to 10 mV/V. The 
default setting is configured to 2 mV/V. Select the value closest to your input range by 
flipping the DIP switches to the appropriate configuration. 
Ex. If you are using a 2 mV/V sensor with the 5 VDC excitation, then you would want to 
use configuration 2 for 1 mV/V. If you are using a 2mV/V sensor with 10 VDC 
excitation then you would want to use configuration 4 for 2 mV/V. 