0 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK IPM650 Product Manual
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Table of Contents 1 2 3 4 5 6 Receiving & Unpacking ................................................................................................................................ - 5 1.1 Unpacking .............................................................................................................................................. - 5 1.2 Storage....................................................
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.6 Auto Reset Timer .............................................................................................................. - 50 6.1.4 Edit Channel .................................................................................................................................. - 51 6.1.4.1 Sensor Configuration ......................................................................................
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.3 Alarm Relay 2 ................................................................................................................................ - 80 6.4.4 Voltage Configuration .................................................................................................................. - 81 6.4.5 Current Value ...........................................................................................
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 1 Receiving & Unpacking 1.1 Unpacking Check all parts to ensure no damage occurred during transit. If you suspect that your product is damaged, contact FUTEK at futek@futek.com for immediate support. 1.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 3 Important Information for IPM650 3.1 Performance The IPM650 is a microcontroller based digital panel meter that monitors the activity of a vast range of sensors. This includes Wheatstone bridge strain gauge and amplified output Torque/ Force/ Pressure/ Displacement sensors.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Built-in shunt calibration features with indicator Lockout feature to prevent inadvertent changes Data logging with up to 21,000 points Supports USB link port Built-in load cell excitation voltage Selectable automatic display shutoff for period of inactivity (up to 15 minutes) Selectable latched or non-latched alarm configuration ASCII Stream Output.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 4 Connector & Wiring Diagram Connectors Amplified Input Description Pin 1 2 3 4 5 6 7 8 Symbol G G 24V OUT 5V OUT -V +V PLEAD PLAG Description Ground Ground 24VDC Output 5VDC Output -V from sensor +V from sensor n/a n/a Strain Gauge Input Pin 1 2 3 4 5 6 Symbol G TEDS -S +S -E +E Description Ground TEDS Data -Signal +Signal -Excitation +Excitation Relays Pin 1 2 3 4 Symbol -R2 +R2 -R1 +R1 Descript
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 5 Features 5.
Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 3.5" 3.5" Screw Terminals Bumpons Tactile Buttons Panel Mount Jacks Black Powder Coated Aluminum Enclosure 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Front Panel Button Function in Menu Mode Enter Function in Normal Mode Displays channel information ▲Arrow Allows the user to toggle between displaying the tare and gross values. This can be used to remove any fixture weights. Back Allows the user to toggle between two different size fonts.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 1 Engineering units can only be converted within its own category.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 5.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 5.3 Main Display FUTEK ADVANCED SENSOR TECH INC. IPM650-PRO MADE IN USA AUTO CALIBRATION IN PROGRESS PLEASE WAIT 11 .... Welcome message will be displayed at power on Auto calibration will be applied when device is turned on. The calibration will minimize any effects of the temperature drift for either zero (offset) or span. During the auto calibration process the key pad is locked out and inaccessible.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK After 10 seconds the following message will be displayed. +2.00001 Peak +2.00000 mV/V +1.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Pressing the display key will change the reading to a larger font. Pressing it again will switch between modes while still in the large font. The tare and gross modes can also be used while in the larger font mode. PK PK +2.00001 +0.00001 01 GROSS 01 TARE mV/V mV/V TR TR +2.00000 +0.00000 01 GROSS 01 TARE mV/V mV/V VY VY +1.99999 -0.00001 01 GROSS mV/V 01 TARE +2.0000 Peak +2.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK The shunt key can be pressed any time after the device has been calibrated and a new sensor profile has been loaded. The simulated value is to be used as a calibration reference. SHUNT will blink on the display. +3.03461 Peak +3.03460 mV/V +1.59484 Valley 01 Gross SHUNT When both SHUNT and HOLD are active on the last row, the display will switch between blinking SHUNT and blinking HOLD.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6 Main Menu Overview EXIS TING CH(S ) S ENS OR CONFIG VIEW CHANNEL DIRECTION NEW CHANNEL UNIT S ELECTION EDIT CHANNEL S ENS OR CAPACITY S AVE CHANGES S ENS ITIVITY (+) DELETE CHANNEL S ENS ITIVITY (-) CALIBRATION DIGITS S ELECT CHANNEL S ELECT MOVING AVERAGE PULS E PER ROTATION S ERIAL NUMBER LIMIT & THD FORCE (MAS S ) TORQUE PRES S URE DIS PLACEMENT mV/V ZERO LOAD (+) FULLS CALE (+) ZERO LOAD (
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Sensor Profile System setting Data Logging Main Menu Output Config interfaces LCD Setting Lock Setting TEDS Data Diagnostic EM1004-B - 19 -
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK SENSOR PROFILE SYSTEM SETTING ► DATA LOGGING OUTPUT CONFIG INTERFACES LCD SETTING ► LOCK SETTING TEDS DATA LOCK SETTINGS TEDS DATA ► DIAGNOSTIC ------------ Press the MENU key to enter the main menu. Use ▲▼keys to select the desired menu option. Press ENTER to navigate to the submenu. The EXIT button can be used at any time to exit any menu or sub menu and return to the display page.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Select “SENSOR PROFILE” from the main menu options and press ENTER. The following options are available: EXISTING CH(S), VIEW CHANNEL, NEW CHANNEL, EDIT CHANNEL, SAVE CHANGESand DELETE CHANNEL. ► EXISTING CH(S) ► EDIT CHANNEL VIEW CHANNEL SAVE CHANGES NEW CHANNEL DELETE CHANNEL EDIT CHANNEL 6.1.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.2 View Channel Scroll to select “VIEW CHANNEL” form the “SENSOR PROFILE” using the▲▼ keys and press ENTER to view: channel number, sensor type, serial number, capacity and engineering unit. If currently more than one channel exists, pressing the arrow ◄►keys on the keypad will cycle through the existing channels.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3 New Channel Scroll to “NEW CHANNEL” from the “SENSOR PROFILE” using the▲▼ keys and press ENTER. Use the ◄► to select the remaining 2 to 15 channels. Press ENTER to set the data for desired channel. Note: The first channel is already set as default and will be displayed on the screen if no other channel has been set.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK This function is not available if lockout option is enabled. See section 6.7 for further information how to enable or disable lock settings. If this device is being used for the first time, it has not been locked yet.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.1 Sensor Configuration Scroll to “SENSOR CONFIG” from “NEW CHANNEL” using ▲▼ keys and press ENTER to set a new channel. “SENSOR CONFIG” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel 1. Use the ◄►keys to select the sensor type as either “FULL BRIDGE”,”VOLTAGE OUTPUT”, or “CURRENT OUTPUT”.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.2 Direction Scroll to “DIRECTION” from “NEW CHANNEL” using the▲▼ keys and press ENTER. “DIRECTION” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel one. Using the ◄►keys on the keypad select either uni-direction or bi-direction.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3 Unit selection DIRECTION ► UNIT SELECTION SENSOR CAPACITY SENSITIVITY(+) Scroll to “UNIT SELECTION” from “NEW CHANNEL” using the ▲▼ keys and press ENTER. It can also be selected from “EDIT CHANNEL” to make modifications on the existing channel except channel one.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3.1 Force (MASS) Scroll to “FORCE (MASS)” from “UNIT SELECTION” using the▲▼ keys and press ENTER. The following options can be selected for force: µg, mg, g, kg, M- tone, dyn, kdyn, Mdyn, N, kN, oz, lbs, klb, ton (US) and ton (UK). Once the desired force unit is selected, the following message will appear.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3.2 Torque Scroll to “TORQUE” from “UNIT SELECTION” using the▲▼ keys and press ENTER. The following options for torque can be selected: g-mm, g-cm, g-m, kg-cm, kg-m, N-mm, N-cm, N-m, kN-m, in-oz, inlb, and ft-lb. Once the desired torque unit is selected, the following message will be shown.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3.3 Pressure Scroll to “PRESSURE” from “UNIT SELECTION” using the▲▼ keys and press ENTER. The following options can be selected for pressure: Pa, kPa, mbar, bar, MPa, kg/cm², atm (standard atmosphere), mmHG, in-H2O, ft-H2O, psi, and kpsi. Once the desired pressure unit is selected, the following message will be shown.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3.4 Displacement Scroll to “DISPLACEMENT” from “UNIT SELECTION” using the▲▼ keys and press ENTER. The following units can be selected for displacement: mm, cm, dm, m, km, in, ft, yd, and mi. Once the desired displacement unit is selected, the following message will be shown.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.3.5 mV/V Scroll to “mV/V” from “UNIT SELECTION” using the ▲▼ keys and press ENTER. Once the desired displacement unit is selected, the following message will be shown.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.4 Sensor Capacity Scroll to “SENSOR CAPACITY” from “NEW CHANNEL” using the▲▼ keys and press ENTER. “SENSOR CAPACITY” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel one. Using the◄► keys move the cursor to the desired placement and use the ▲▼keys to select the desired number or decimal position. The capacity can be selected from 0.00001 to 9999999.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.5 Sensitivity (+) SENSOR CONFIG SELECT SENSOR ► SENSITIVITY(+) mV/V OUTPUT USING ARROW KEYS UNIT SELECTION [2.00000] SENSOR CAPACITY Scroll to “SENSITIVITY (+)” from “NEW CHANNEL” using the▲▼ keys and press ENTER to set the sensitivity for a new channel. “SENSITIVITY” can also be selected from “EDIT CHANNEL” to make modifications on the existing channel except channel one.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Note: For voltage sensor device asks for voltage output. 6.1.3.6 Sensitivity (-) SENSOR CONFIG SELECT SENSOR ► SENSITIVITY(-) mV/V OUTPUT USING ARROW KEYS UNIT SELECTION [2.00000] SENSOR CAPACITY Scroll to “SENSITIVITY (-)” from “NEW CHANNEL” using the ▲▼ keys and press ENTER to set the sensitivity for a new channel.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK The following message will be seen if the sensor sensitivity is the same as current profile.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.7 Calibration Scroll to “CALIBRATION” from “NEW CHANNEL” using ▲▼ keys and press ENTER to set it for the new channel. “CALIBRATION” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel one. Note: This function is dedicated for precise calibration (live calibration), however if the user would not like to apply the load, in the sensitivity selection (6.1.3.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.7.2 Full Scale (+) Scroll to “FULL SCALE (+)” from “CALIBRATION” using the ▲▼ keys and press ENTER. APPLY OPERATION FULL SCALE(+) WAS EXECUTED CALIBRATION SUCCESSFULLY THEN PRESS ENTER (HIT ENTER/BACK) If the channel “DIRECTION” is defined as “BI-DIRECTION” it will be prompt to apply the full scale calibration (-) as well as zero load (-) for the reverse direction. 6.1.3.7.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.8 Serial number Scroll to “SERIAL NUMBER” from “NEW CHANNEL” using the ▲▼ keys and press ENTER. “SERIAL NUMBER” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel one. CALIBRATION PULSE/ROTATE ► SERIAL NUMBER LIMIT & THD Using the◄► keys, the cursor can be moved among the seven digits and use the ▲▼ keys to select the desired serial number.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9 Limit & THD Scroll to LIMIT & THD” from “NEW CHANNEL” using the▲▼ keys and press ENTER.” LIMIT & THD” can also be selected from “EDIT CHANNEL” to make modifications on the existing channels except channel one.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.1 First Peak THD Scroll to “FIRST PK THD” from “LIMIT & THD” using the▲▼ keys and press ENTER. ► FIRST PK (THD) FIRST VY (THD) MIN / MAX DIFF ALARM LIMIT HI Use the arrow keys to move the cursor among the digits and select the desired number for the First Peak. The sign value can also be changed from positive to negative in case a bi-directional sensor is configured.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.2 First Valley THD Scroll to “FIRST VY (THD)” from “LIMIT & THD” using the ▲▼ keys and press ENTER. FIRST PK (THD) ► FIRST VY (THD) MIN / MAX DIFF ALARM LIMIT HI Use the arrow keys to move the cursor among the digits and select the desired number for the First Valley. The sign value can also be changed from positive to negative in case a bi-directional sensor is configured.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK FP Unit FP THD 80% THD FV THD 10% t FV Figure No.1 Example of One Directional Sensor FP Unit 50% FP THD THD t FV THD -50% FV Figure No.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.3 MIN/MAX Differentiation Scroll to “MIN / MAX DIFF” from “LIMIT & THD” using the▲▼ keys and press ENTER. FIRST PK (THD) FIRST VY (THD) ► MIN / MAX DIFF ALARM LIMIT HI Use the arrow keys to move the cursor among the digits and select the desired number for the Min/Max Differentiations. The sign value can also be changed from positive to negative in case a bi-directional sensor is configured.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK A X% Max/Min Diff B Figure No.3 Max/ Min Differentiation Caution should be taken in defining X, for the Max/Min Diff. For example: Consider the following data at points A and B: A: 100 IL B: 90IL If X is considered less than 10%, A would not be considered as first peak, or B would not be considered as first valley.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.4 Alarm Limit High Scroll to “ALARM LIMIT HI” from “LIMIT(S) & THD(S)” using the▲▼ keys and press ENTER. FIRST VY THD MIN / MAX DIFF ► ALARM LIMIT HI ALARM LIMIT LO Use the arrow keys to move the cursor among the digits and select the desired number for the Alarm Limit High. The sign value can also be changed from positive to negative in case a bi-directional sensor is configured.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.5 Alarm Limit Low Scroll to “ALARM LIMIT LO” from “LIMIT & THD” using the ▲▼ keys and press ENTER. MIN/MAX DIFF A ALARM LIMIT HI ► ALARM LIMIT LO AUTO RST (SEC) Use the arrow keys to move the cursor among the digits and select the desired number for the Alarm Limit Low. The sign value can also be changed from positive to negative in case a bi-directional sensor is configured.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Alarm Area Unit 80% Non Alarm Area 10% t Alarm Area Figure No.4 Example of One Directional Sensor Alarm Area Unit 50% Non Alarm Area -50% t Alarm Area Figure No.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.3.9.6 Auto Reset Timer Scroll to “AUTO RST (SEC)” from “LIMIT & THD” using the▲▼ keys and press ENTER. ALARM LIMIT HI ALARM LIMIT LO ► AUTO RST (SEC) --------Use the arrow keys to move the cursor among the digits and select the desired number. SELECT DESIRED VALUE USING ARROW KEYS [0010] A confirmation message will be displayed confirming operation was successfully executed.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.4 Edit Channel Scroll to “EDIT CHANNEL” from “SENSOR PROFILE” using the▲▼ keys and press ENTER.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Use the◄► to select the channel to edit. SELECT DESIRED CHANNEL NUMBER USING ◄,► KEYS [02] All the features that are accessible in the “New CHANNEL” option, applies to the Edit Channel option as well.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.4.1 Sensor Configuration See the details mentioned in section 6.1.3.1. 6.1.4.2 Direction See the details mentioned in section 6.1.3.2. 6.1.4.3 Unit selection See the details mentioned in section 6.1.3.3. 6.1.4.3.1 Force (MASS) See the details mentioned in section 6.1.3.3.1. 6.1.4.3.2 Torque See the details mentioned in section 6.1.3.3.2. 6.1.4.3.3 Pressure See the details mentioned in section 6.1.3.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.4.7.3 Zero Load (-) See the details mentioned in section 6.1.3.7.3 6.1.4.7.4 Full Scale (-) See the details mentioned in section 6.1.3.7.4 6.1.4.8 Serial number See the details mentioned in section 6.1.3.8. 6.1.4.9 Limit & THD See the details mentioned in section 6.1.3.9. 6.1.4.9.1 First Peak (THD) See the details mentioned in section 6.1.3.9.1. 6.1.4.9.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.5 Save Changes Scroll to “SAVE CHANGES” from “SENSOR PROFILE” using the▲▼ keys and press ENTER to save the modifications that are made under the Edit Channel or New Channel menu. VIEW CHANNEL NEW CHANNEL EDIT CHANNEL ► SAVE CHANGES The Save Changes option is not available if the lockout option is enabled. See section 6.7 for further information on lock settings.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Anytime that new sensor sensitivity is set or the sampling rate is changed, the device may do an auto calibration after it loads the new profile.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.1.6 Delete Channel Scroll to “DELETE CHANNEL” from “SENSOR PROFILE” using the▲▼ keys and press ENTER. Use the ◄►keys to select the channel to delete. Choose Accept or Cancel to confirm or cancel the operation. Note: Any channel other than the first channel can be deleted.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK WARNING! SELECTED PROFILE MAY BE LOST ACCEPT CANCEL A message confirming the desired operation was successfully executed will be displayed.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK LOADING PROFILE AUTO CALIBRATION IN PROGRESS IN PROGRESS PLEASE WAIT 30 PLEASE ……… WAIT 15 BACK can be applied to cancel or ENTER can be used to delete the selected channel.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.1 Digit Select Scroll to “DIGIT SELECT” from “SYSTEM SETTING” using the▲▼ keys and press ENTER. ► DIGIT SELECT CHANNEL SELECT MOVING AVERAGE SAMPLING RATE Use the◄► keys to select the number of digits from 3 to 6. SELECT DESIRED DISPLAY DIGITS USING ◄,► KEYS [6] Once the desired number of digits has been selected, a prompt will appear prompting to set the value as a default value.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.2 Channel Select Scroll to “CHANNEL SELECT” from “SYSTEM SETTING” using the▲▼ keys and press ENTER. Use the ◄► keys to select the desired number. If no other channels have been added, only channel 01 can be selected, which is the default channel.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.3 Moving Average Scroll to “MOVING AVERAGE” from “SYSTEM SETTING” using the▲▼ keys and press ENTER. Using the◄► keys: disable, 2, 4, 8, 16, 32, 64, or 128 can be selected.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK This is known as a moving average because the average at each kth instant is based on the most recent set of n values. In other words, it is a moving window of n values that are used to calculate the average of the data sequence (see Figure 1). Figure 1: Moving Average of n Data Points The averaging process can be improved if the calculations can be performed in a recursive fashion.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.4 Sampling Rate Scroll to “SAMPLING RATE” from “SYSTEM SETTING” using the ▲▼ keys and press ENTER.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.5 Peak/ Valley Scroll to “PEAK/ VALLEY” from “SYSTEM SETTING” using the▲▼ keys and press ENTER. CHANNEL SELECT MOVING AVERAGE SAMPLING RATE ► PEAK / VALLEY Using the◄►keys select to display the First Peak/Valley or Hold Peak/Valley values. SELECT DESIRED CONFIGURATION USING ◄,► KEYS [HOLD] A prompt will be shown to set this value as a default setting.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK First peak and first valley will be captured based on the specified threshold values defined in the sensor profile. Refer to sections (6.1.3.10.1), (6.1.3.10.2), and (6.1.3.10.3) for more details. Valley is the maximum absolute value in the negative direction.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.6 Auto Reset Scroll to “AUTO RESET” from “SYSTEM SETTING” using the ▲▼ and press ENTER. Auto Reset will reset the peak and valley values when the timer is expired. The time, in seconds, is defined in the sensor profile. Refer to section (6.1.3.9.6) for more detail. SAMPLING RATE PEAK / VALLEY ► AUTO RESET ALARM CONFIG Using the◄►keys select to enable or disable the Auto Reset function.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.7 Alarm Configuration Scroll to “ALARM CONFIGURATION” in “SYSTEM SETTING” using the▲▼ keys and press ENTER. The Alarm threshold levels are set in the sensor profile (Refer to section 6.1.3.9), Refer to the drawings on page 44 for examples regarding alarm THD levels. SAMPLING RATE PEAK / VALLEY AUTO RESET ► ALARM CONFIG Using the◄►keys set the Alarm Configuration to Latched or Non-Latched.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.2.8 Alarm Activity Scroll to “ALARM ACTIVITY” from “SYSTEM SETTING” using the▲▼ keys and press ENTER. AUTO RESET ALARM CONFIG ► ALARM ACTIVITY ------------Using the◄►keys different alarm activities including System Alarm or Relay Alarm can be either enabled or disabled. SELECT DESIRED ACTIVITY USING ◄,► KEYS [DISABLE] A prompt will be shown to set this value as a default setting.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.3 Data Logging Data Logging Rate Data Logging Duration Action SENSOR PROFILE SYSTEM SETTING ► DATA LOGGING OUTPUT CONFIG This menu option allows the user to log the data. Data logging is a useful feature to monitor the performance of sensor during a specific time.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.3.1 Logging Rate Scroll to “LOGGING RATE” from “DATA LOGGING” using the▲▼ keys and press ENTER. Using the◄► keys select either HALF SPEED, FULL SPEED. This setting can be set as a default setting.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.3.2 Duration (SEC) Scroll to “DURATION” from “DATA LOGGING” using the▲▼ keys and press ENTER. Use the◄► and the▲▼ keys to select the desired duration in seconds; therefore, from 0 to 9999 seconds can be selected.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Although any other numbers can be selected, after the actual maximum duration time, the internal buffer will hit full capacity and the following message will appear: DATA LOGGING PROCESS IS COMPLETED SUCCESSFULLY EM1004-B - 74 -
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.3.3 Action Scroll to “ACTION” from “DATA LOGGING” using the▲▼ keys and press ENTER. Using the ◄►keys, the data logging session either can be started or stopped. This sub menu allows the user to Stop or Start data logging process. As soon as the Start option is selected, the data logging process starts.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK It is important to know that IPM650 is equipped by a high speed Processor. This processor is responsible to control every single feature of device and dedicates the specific time frame for internal operation and Interfaces.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.1 Digital ASCII Scroll to “DIGITAL ASCII” from “OUTPUT CONFIG” using the▲▼ keys and press ENTER. ASCII output consists of four rows of data.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.2 Alarm Relay 1 Alarm relay 1 is a solid state relay with a voltage / current rating of 110V/100mA. Alarm relay 1 has been mapped to alarm high. Scroll to “ALARM RELAY1” from “OUTPUT CONFIG” using the▲▼ keys and press ENTER. Use the◄►keys to set the alarm relay 1 either to be normally open or normally close.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.3 Alarm Relay 2 Alarm relay 2 is a solid state relay with a voltage / current rating of 110V/100mA. Alarm relay 2 has been mapped to alarm low. Scroll to “ALARM RELAY2” from “OUTPUT CONFIG” using the▲▼ keys and press ENTER. Use the◄►keys to set the alarm relay 2 either to be normally open or normally close.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.4 Voltage Configuration Scroll to “VOLTAGE CONFIG” from “OUTPUT CONFIG” using the▲▼ keys and press ENTER. ALARM RELAY1 ALARM RELAY2 ► VOLTAGE CONFIG CURRENT VALUE Use the◄►keys to select uni-polar which is 0 to 5 volts (negative full scale is mapped to 0 VDC, zero load is mapped to 2.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.5 Current Value Scroll to “CURRENT VALUE” from “OUTPUT CONFIGURATION” using the▲▼ keys and press ENTER to define the current value. ALARM RELAY2 VOLTAGE CONFIG ► CURRENT VALUE CURRENT CONFIG Use the◄► keys and press ENTER to select any of the following Ranges: 0-20 mA, 4-20 mA, 0-25 mA or 5-25 mA.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.4.6 Current Configuration Scroll to “CURRENT CONFIG” from “OUTPUT CONFIG” using the▲▼ keys and press ENTER. Use the◄►keys to select either UNIDIRECTIONAL or BIDIRECTIONAL For a Bi-Directional configuration the, negative full scale output from the sensor is mapped to the lowest value of current.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.1 USB Output Scroll to “USB OUTPUT” from “INTERFACE” using the▲▼ keys and press ENTER. ► USB OUTPUT¾ ASCII OUTPUT RELAY1 OUTPUT RELAY2 OUTPUT Use the◄►keys to either enable or disable the USB output activity. SELECT DESIRED ACTIVITY USING ◄,► KEYS [ENABLE] A prompt will be shown to set this value as a default setting.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.2 ASCII output Scroll to “ASCII OUTPUT” from “INTERFACE” using the▲▼ keys and press ENTER. USB OUTPUT¾ ► ASCII OUTPUT RELAY1 OUTPUT RELAY2 OUTPUT Use the◄► keys to either enable or disable the ASCII output. The termination character configuration should be set in the Digital ASCII menu within the Output Config menu. Refer to section (6.4.1) for more details.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.3 Relay 1 Output Scroll to “RELAY1 OUTPUT” from “INTERFACE” using the▲▼ keys and press ENTER. USB OUTPUT¾ ASCII OUTPUT ► RELAY1 OUTPUT RELAY2 OUTPUT Use the◄► keys to enable or disable Relay 1 output. The configuration of the Relay 1 Output should be set under the output configuration menu. Refer to section (6.4.2) for more details.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.4 Relay 2 Output Scroll to “RELAY OUTPUT2” from “INTERFACE” using the▲▼ keys and press ENTER. ASCII OUTPUT RELAY1 OUTPUT ► RELAY2 OUTPUT VOLTAGE OUTPUT Use the ◄►keys to either disable or enable the Relay 2. The configuration of the Relay 2 Output should be set under the output config menu. Refer to section (6.4.3) for more details.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.5 Voltage Output Scroll to “VOLTAGE OUTPUT” from “INTERFACE” using the▲▼ keys and press ENTER. RELAY1 OUTPUT RELAY2 OUTPUT ► VOLTAGE OUTPUT CURRENT OUTPUT Use the◄► keys to either disable or enable the Voltage Output activity. The configuration of the Voltage Output should be set under the output config menu. Refer to section (6.4.4) for more details.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.6 Current Output Scroll to “CURRENT OUTPUT” from “INTERFACE” options using the▲▼ keys and press ENTER. RELAY2 OUTPUT VOLTAGE OUTPUT ► CURRENT OUTPUT POWER OUTPUT Use the ◄►keys on to either disable or enable the Current Output. The configuration of the Current Output should be set under the output config menu. Refer to section (6.4.5) for more details.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.5.7 Power Output Scroll to “POWER OUTPUT” from “INTERFACE” options using the▲▼ keys and press ENTER. VOLTAGE OUTPUT CURRENT OUTPUT ► POWER OUTPUT BRIDGE INPUT Use the◄►keys to either disable or enable the Power Output.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Important notice: Bridge sensors are internally connected to input number 1, current sensors are connected to input number 2 and voltage sensors are connected to input number 3. Enabling a different type of input can result in incorrect measurements. 6.5.8 Bridge Input Scroll to “BRIDGE INPUT” from “INTERFACE” using the▲▼ keys and press ENTER.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Whenever the Bridge Input is enabled a check mark showing that this feature is enabled will be seen on the display. Important notice: Bridge sensors are internally connected to input number 1, current sensors are connected to input number 2 and voltage sensors are connected to input number 3. Enabling a different type of input can result in incorrect measurements. 6.5.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK WARNING! EXISTING SETTING MAY BE CHANGED ACCEPT CANCEL Whenever the Voltage Input is enabled a check mark showing that this feature is enabled will be seen on the display. Important notice: Bridge sensors are internally connected to input number 1, current sensors are connected to input number 2 and voltage sensors are connected to input number 3.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Depending on the type of sensor defined in the current sensor profile, a warning message will be displayed when attempting to enable an input type which is not matched with the sensor configuration.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.6.1 Contrast Scroll to “CONTRAST” from “LCD SETTING” using the▲▼ keys and press ENTER. ► CONTRAST BRIGHTNESS AUTO LCD OFF ----------Use the◄►keys to select the desired LCD contrast level. SELECT DESIRED CONTRAST USING ◄ ,► KEYS ≡≡≡≡≡≡≡ A prompt will be shown to set this value as a default setting.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.6.2 Brightness Scroll to “BRIGHTNESS” from “LCD SETTING “using the▲▼ keys and press ENTER. CONTRAST ► BRIGHTNESS AUTO LCD OFF ----------Use the◄► keys to select the desired LCD brightness level. SELECT DESIRED BRIGHTNESS USING ◄ ,► KEYS ≡≡≡≡≡≡≡ The desired Brightness can be saved as default.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.6.3 Auto LCD off Scroll to “AUTO LCD OFF” from the “LCD SETTING” using the▲▼ keys and press ENTER. CONTRAST BRIGHTNESS ► AUTO LCD OFF ----------Use the ◄► keys to Enable or Disable the Auto LCD Off feature. The Auto LCD Off timer can be set for 1-15min, after which the LCD will turn off if left unattended.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.7 Lock Settings Lockout profile Lock Setting Unlock Profile Change Password INTERFACES LCD SETTING ► LOCK SETTING TEDS DATA The Lock Setting menu is used to control access to the Sensor Profile settings and data information that have been stored in internal non-volatile memory (6.1.3.x and 6.1.4.x).
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.7.1 Lockout Profile Scroll to “LOCKOUT PROFILE” from “LOCK SETTING” using the▲▼ keys and press ENTER. Once pressed ENTER, this feature will lockout the Edit, New, Save, and Delete of SENSOR PROFILE menu. However the Existing Channel(s) and the View Channel(s) options are still accessible. ► LOCKOUT PROFILE UNLOCK PROFILE CHANGE PASSWORD ----------Enter the current password to lockout the sensor profile.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.7.2 Unlock profile Scroll to “UNLOCK PROFILE” from “LOCK SETTING” using the▲▼ and press ENTER. Once pressed the ENTER, this feature unlocks the sensor profile menu and full access to sensor profile will be gained. LOCKOUT PROFILE ► UNLOCK PROFILE CHANGE PASSWORD ----------Enter the current password to unlock the sensor profile.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.7.3 Change Password Scroll to “CHANGE PASSWORD” from “LOCK SETTING” using the▲▼ keys and press ENTER. LOCKOUT PROFILE UNLOCK PROFILE ► CHANGE PASSWORD ----------Use the◄► and ▲▼ keys to select the current password. ENTER CURRENT PASSWORD USING ARROW KEYS [0000000] The current password shall be entered in order to be able to change the password.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Use the ◄► and ▲▼ keys to set the desired password. Any combination of seven digits can be chosen. Press ENTER to save the new password. SELECT DESIRED PASSWORD USING ARROW KEYS [0000000] The following success message will appear on the screen once the desired password has been entered.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.8 TEDS Data TEDS Device TEDS Page TEDS Data Load Data Auto Detection LCD SETTING LOCK SETTING ► TEDS DATA DIAGNOSTIC The mentioned IEEE 1451.4 standard defines a collection of templates for common class transducer. The templates provide the means for the measurement system to convert the binary data stored on a smart TEDS sensor EEPROM.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.8.1 TEDS Device Scroll to “TEDS DEVICE” from “TEDS DATA” using the▲▼ keys and press ENTER. ►TEDS DEVICE TEDS PAGE LOAD DATA AUTO DETECTION If a TEDS device is connected, the TEDS information will be displayed as shown below: DEVICE: DS2433 MEMORY: 4K BITS TEDS SERIAL NO: 000000D45CA1H If the TEDS DEVICE is selected without a connected TEDS device, an error message will be displayed.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.8.2 TEDS Page Scroll “TEDS PAGE” from “TEDS DATA” using THE▲▼ keys and press ENTER. A total of sixteen pages (0 – 15) can be selected TEDS DEVICE ► TEDS PAGE LOAD DATA AUTO DETECTION 03E8042000003753 0000000000148400 8000000002412000 2316C0DA28EC8F5C If TEDS PAGE is selected, without a connected TEDS device, an error message will be displayed.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.8.3 Load Data Scroll to “LOAD DATA” from “TEDS DATA” using the▲▼ keys and press ENTER. TEDS DEVICE TEDS PAGE ► LOAD DATA AUTO DETECTION The load data option loads the new profile, based on the TEDS data, and performs an auto calibration. LOADING PROFILE AUTO CALIBRATION IN PROGRESS IN PROGRESS PLEASE WAIT TEMPLATE SERIAL 30 PLEASE WAIT ……… 15 33 1000 +10.0000 lbs +2.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.8.4 Auto Detection Scroll to “AUTO DETECTION” from “TEDS DATA” using the▲▼ keys and press ENTER. If “AUTO DETECTION” is enabled, the IPM650 will search for a TEDS device during power up. If TEDS information exists, its information will be loaded; otherwise the IPM650 will load the active channel. If “AUTO DETECTION” is disabled, the IPM650 will load the active channel during power up.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 6.9.1 Internal or External Scroll to “DIAGNOSTIC” using the▲▼keys and press ENTER. Use the◄► keys to select either INTERNAL or EXTERNAL diagnostic. SELECT DESIRED AVDD1 : 4.999V DIAGNOSTIC AVDD2 : 5.200V USING ◄,► KEYS DVDD : 5.001V [INTERNAL] VBAT : 8.500V The internal diagnostic shows the analog VDD1, analog VDD2, digital VDD, and VBAT values.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 7 Appendix A (List of Probable Errors) Error No.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 8 Appendix B (List of Messages) Message No. Welcome Message Delete Message Delete Confirmation Interface Warning Unsaved Data Warning Success Message Confirmation Message Edit/ delete Warning (+)No Load Calibration Message Description on the LCD FUTEK ADVANCED SENSOR TECH INC.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK (+)Full Load Calibration (-)Full Load Calibration (-)No Load Calibration Unavailable Feature Incorrect password Warning Data Logging Confirmation APPLY FULL SCALE(+) CALIBRATION THEN PRESS ENTER APPLY FULL SCALE(-) CALIBRATION THEN PRESS ENTER APPLY ZERO LOAD(-) CALIBRATION THEN PRESS ENTER SELECTED FEATURE IS NOT AVAILABLE AT THIS TIME HIT ENTER / BACK INVALID PASSWORD HAS BEEN ENTERED TRY AGAIN LATER!
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 9 Appendix C (Device Specifications) Main Control Processor Unit (MCPU) Part #: MSC1214 Internal ADC Resolution: 24 bits Sampling Rate: 60 SPS Application: Diagnostic mode DAC Resolution: * DAC Accuracy: * DAC Non linearity:* Application:* 16 bits 0.02% of FSR (Factory Calibrated) 0.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Liquid Crystal Display (LCD) Part #: Brand: Number of Characters (Width): No. of Lines (Lines): Backlight Type & Color: Fluid Type, Image (Positive or Negative), & LCD glass Color: Polarizer Film Type, Wide (WT) Temperature Range, & Viewing Angle (O’clock): Character Set (CGROM): Controller: CFAH1604B-TMI-ET Crystalfontz America, Inc.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK Default Password: Alternative Password: 0000000 Device’s Serial Number Instrument Weight: Dimensions: 1.9 lbs. (862 g) 3.8” (L) x 3.8” (W) x 6.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 10 Appendix D TEDS IEEE 1451.4 Introduction 10.1 What is TEDS? Plug and play sensor hardware and software makes configuring a smart TEDS sensor as easy as plugging a mouse into a PC. Efficiency and productivity has greatly improved due to the elimination of manual sensor configuration. 10.2 Basic Concept TEDS is at the heart of the new universally accepted IEEE1451.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK 11 Appendix E (System Performance) Specifications: Parameter Output Data Rate No Missing Codes Resolution RMS Noise and Output Data Rates Integral Nonlinearity Offset Error Offset Error Drift vs. Temperature Offset Error drift vs. Time Gain Error Gain Drift vs. Temperature Gain Drift vs.
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK RMS Noise (µV): Output data Settling Time S≤500mV/V* S≤7.5mV/V Rate (Hz) (ms) 5 818 0.256 0.640 10 400 0.330 0.896 25 160 0.544 1.344 30 133.4 0.615 1.472 50 80 0.900 2.112 60 66.7 0.970 2.304 100 40 1.161 2.816 200 20 1.567 3.968 400 10 2.067 5.440 600 6.66 2.400 6.656 800 5 2.734 7.936 960 4.17 3.000 8.960 1200 3.33 3.334 9.856 1600 2.5 3.8889 11.264 2400 1.67 5.000 14.080 4800 0.83 14.300 24.