This drawing is submitted solely for the information and
exclusive use of the original addressee. It is not to be divulged
in whole or in part, by any firm or individual without written
permission from FUTEK
Drawing Number: F3131
RECESSED
PLATE
RETAIN WITH
SET SCREW
CAUTIONS
MOUNTING
MAINTAIN CLEAN
SURFACE
SUPPORT ON THE
OUTER RING ONLY.
1) DO NOT PULL ON OR CARRY SENSOR BY CABLE.
2) BE CAREFUL NOT TO APPLY EXCESSIVE FORCE ON
OR MOUNT ANYTHING IN CONTACT OF THE
DIAPHRAGM AREA.
3) ANY TAMPERING OR REMOVAL OF COVER, CABLE
AND/ OR CONNECTOR WILL VOID WARRANTY
4) ALWAYS HAVE SENSOR PLUGGED IN DURING
INSTALLATION TO MONITOR OUTPUT TO AVOID
PERMANENT ZERO SHIFT AND OVERLOAD.
LOAD SHOULD BE
APPLIED IN LINE
LOAD SHOULD BE
APPLIED ON THE
CENTER OF THE
BUTTON
SECURE FROM
OUTER LIP
LOAD BUTTONS ARE USED IN
THE MONITORING OF FORCE
APPLIED IN PRESS TYPE
APPLICATIONS.
FIXTURE MUST NOT
CONTACT DIAPHRAGM
AREA
FIXTURE LOAD SURFACE
MUST BE IN LINE WITH THE
SENSOR
SENSOR CAN NOT BE
USED IN TENSION.
SIDE LOAD MAY DAMAGE
THE SENSOR
THREADED LOAD PROBE TYPESTANDARD
FLAT AND EVEN SURFACE
REQUIRED
1 2
43
65
7 8
FURTHER MOUNTING SUGGESTIONS
DIAPHRAGM
DIAPHRAGM
DIAPHRAGM
MOUNT IN
C'BORE HOLE
MOUNT USING
THREAD HOLES
SENSOR MAY BE
USED TO
MONITOR
FORCES
IN AN ARBOR
PRESS
DIAPHRAGM