The next Tab or Window under the Information Tabs section of the Calibration / Verification interface is the Initial Tab, where you will do 
the Initial setup for your test: Record Environmental conditons, perform Shunt Calibration with either the built-in Shunt values,or with your 
own, external shunt resistors, and also take and record Bridge Resistance readings.
The first section of the Initial tab is the Environmental 
Conditions section; here you can enter and record the relative 
humidity levels and leakage factor for your test, as well as obtain 
and record the temperature in either Fahrenheit or Celsius from 
your VCal™ temperature probe.
Next is the Shunt Calibration section; here you can use VCal™’s 
built in Shunt values or select External and connect your own 
shunt resistors to VCal™, and obtain and record the Shunt 
Calibration readings for your test.
The final section of the Initial tab is the Bridge Impedance 
section; here you can obtain and record the bridgeresistance 
for the sensor you are calibrating by clicking the Take Reading 
Button
The Next Tab icon will move you to the next window, in this case, to 
the Setup Tab . . .
The Setup Tab is where you can define the Testing Mode,
your Test Options, the Test Parameters, and the Test 
Conditions for your test.
Under Testing Modes you can choose between Normal, 
Extrapolation, and Actual Modes of Testing. Descriptions 
of these Modes will be given in the Calibration/Verification 
Instructions section of this Manual. (pg. 42)
Under Test Options you can choose between either Manual 
Entry, or Scaled Testing. More detail will be given on these 
options in the Calibration/Verification Instructions section of 
this Manual. (pg. 42)
Under Test Parameters you can define such parameters for your 
test as Loading Points, Averaging, Decimal Points, and Target 
Load Warning Tolerances. The Scaled Testing Decimal 
Points textbox allows you to select output decimal points when 
performing a scaled test. More details will be given on setting 
these parameters in the Calibration/Verification Instructions 
section of this Manual. (pg. 43)
Under Test Conditions you can define whether your test is 
performed with VCal™ automatically taking readings at your 
pre-determined load points, or whether you wish to read your 
data from an RS232 device, and also whether you wish to test for 
Linearity or Linearity / Hysteresis. There is also a button for 
Beginning your test. (pg. 43) 