Introduction To Performing A VCal™ Calibration / Verification
You are nearly ready to do some calibrating, however; before we step through some Calibration Examples, there are several items which require 
some additional explanation. Testing Modes, Testing Options, Testing Parameters and Test Conditions are all items which really need 
to be explained a little before we proceed. All of these can be found under the Setup tab. We will also talk some more about the three types of 
testing you can do in the Calibration / Verification interface; the Main Test, the Time Test, and the Signature Test.
Testing Modes
Under the Testing Modes section of the Setup tab you have a choice of whether you wish to use Normal, Extrapolation, or Actual Modes, 
when you perform your test. Let’s define what these calibration modes actually are.
Normal Mode Calibration: This is your default calibration mode. In this type of calibration you load your sensor to as close as possible to the 
load points indicated in the Load column of the Test Grid window, pressing Enter at each load point. Each time you press Enter, VCal™ takes an 
output reading and calculates the non-linearity (and hysteresis if selected), based on the deviation from the nominal or ideal value.
Extrapolation Mode Calibration: In this calibration mode you load your sensor to a point somewhat close to the indicated load point
(max+/-10% of load), and press Enter, and VCal™ takes a reading at the point you actually hit Enter at and extrapolates the output at the the 
load value indicated in the Test Grid. It calculates the non-linearity (and hysteresis if selected) based on the deviation from the nominal. For 
Example; If you pressed Enter at 79.3 lbs load, and your output was 0.5335 mV/V, but the load indicated in the Test Grid is 80 lbs, then; 80.0 
lbs X (0.5335/79.3) = 0.5402 mV/V (this would be VCal™’s output reading in Extrapolation Mode under the test conditons described.
  Note: Extrapapolation Mode only works when you have a calibrated reference cell connected as well as your test cell, and all 
  channels must be designated correctly for the calculations to work correctly.
Actual Mode Calibration: In this mode you take readings at any pattern of points you desire, pressing Enter at each point, and VCal™ takes 
a reading at that point, and changes the indicated loads in the Test Grid to reflect the points at which you pressed Enter. VCal™ then gives 
an output reading at that point, and calculates the non-linearity based on the deviation from the new nominal value. For Example; If the load 
indicated on VCal™’s Test Grid is 80 lbs and you press Enter when there is only 79.3 lbs of load applied, then VCal™ will change the value 
indicated in the Test Grid to 79.3 lbs and will display the output value that corresponds to 79.3 lbs of load.
  Note: Because hysteresis requires exactly matching ascending as well as descending load points, it is not available in this mode of 
  calibration.
Testing Options
Manual Entry: Manual Entry allows you to enter data from a certificate or a different data acquisition system, through the keyboard, directly into 
the Test Grid. The use of Manual Entry actually prevents the program from obtaining data from the VCal™ module. With Manual Entry you 
can enter the Load and Output values from your test, and VCal™ will caluclate the non-linearity (and hysteresis if you selected it) as soon as 
you press Enter in the final zero output box in the Test Grid.
Scaled testing: Scaled Testing is designed to be used when data is being entered from an outside source which has already been scaled to a 
known unit, such as a pre-calibrated Sensor-Display system (Futek D500 series). If you are using a sensor that is giving you a raw output in Volts 
or mV/V you cannot use Scaled Testing, because your scaling factor is unknown at this time. Using Scaled Testing changes the way in which errors 
are calculated. Below are a couple of examples which illustrate the procedures used to calculate errors in both Scaled and Non-Scaled Testing:
  Non-Scaled Example: (Output in mV/V)        Scaled Testing: (known output in lbs, kg, etc.)
    0  0.0000              0  0.0
    20  0.3995              20  19.8
 LOAD  40  0.8000   OUTPUT       LOAD  40  39.9   OUTPUT
    60  1.2000              60  60.0
    80  .6000              80  79.7
    100  2.0000              100  101.3       
 In this case the error is a percentage of full-scale, for example
 the error at 20 lbs is ((0.4000 - 0.3995) / 2.0000) * 100
 = 0.025% of full-scale. Note that 0.4000 is the ideal output
 at a load of 20 lbs.
 In this case the error is a percentage of load, for example; the error                    
 at 20 lbs is ((20 - 19.8) / 20) * 100 = 1% of the Load.