Users Manual

Operating Manual Smart Weather Sensor
108 G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany
19.6.5.7 Buffer Assignment Basic Data Set WS501-UMB, WS502-UMB, WS503-UMB,
WS504-UMB, WS510-UMB
Device configured for measurement values in metric units:
Measurement value
UMB
Channel
Min
Max
Unit
Buffer ‘0’
Air temperature (act)
100
-50.0
60.0
°C
Rel. Humidity (act)
200
0.0
100.0
%
Rel. Air Pressure
305
300.0
1200.0
hPa
Wind Speed (avg)
460
0.0
75.0
m/s
Wind Speed (max)
440
0.0
75.0
m/s
Buffer ‘1’
Wind Direction (vct)
580
0.0
359.9
°
Global Radiation (act)
900
0.0
2000.0 *)
W/m²
Dew Point (act)
110
-50.0
60.0
°C
Wind Chill Temperature (act)
111
-60.0
70.0
°C
Buffer ‘2’
Wind Speed (act)
400
0.0
75.0
m/s
Wind Speed (min)
420
0.0
75.0
m/s
Wind Speed (vct)
480
0.0
75.0
m/s
Wind Quality
805
0.0
100.0
%
Buffer ‘3’
Wind Direction (act)
500
0.0
359.9
°
Wet Bulb Temperature (act)
114
-50.0
60.0
°C
Specific Enthalpy (act)
215
-100.0
1000.0
kJ/kg
Buffer ‘4’
Global Radiation (min)
920
0.0
2000.0 *)
W/m²
Global Radiation (max)
940
0.0
2000.0 *)
W/m²
Global Radiation (avg)
960
0.0
2000.0 *)
W/m²
*) Until device version .49 / .212 / Firmware v5.6: 1400.0 W/m
2
Example: Request Buffer ‘0’
0D0!
0+13.5+85.7+2.5+3.7<CR><LF>
Air Temperature 13.5°C, rel. Humidity 85.7%, average wind speed 2.5m/s, peak wind speed 3.7m/s