User's Guide

Dexcom G6 System User Guide
306
Appendix F: Technical Information
Table 5-B. Hypoglycemic Alert and Detection Rate Evaluations
(Pediatrics*, n=165
1
)
Hypoglycemic
Alert Level
(mg/dL)
Alerts Detections
# of
alerts
(n)
True
Alert
Rate (%)
False
Alert
Rate (%)
# of
events
(n)
Correct
Detection
Rate (%)
Missed
Detection
Rate (%)
55 358 31.6 68.4 66 68.2 31.8
60 521 44.1 55.9 119 73.1 26.9
70 1,054 68.0 32.0 369 81.6 18.4
80 1,794 80.5 19.5 671 88.1 11.9
90 2,746 86.3 13.7 1,030 92.8 7.2
* Includes pediatric subjects 6-17 years of age; no YSI measurements were taken for
pediatric subjects 2-5 years of age.
1
All subjects were considered in the analysis; however, not all subjects experienced
hypo event
LBL014003 Rev XXX Artwork, G6 CGM System User Guide US.indd 306 2/5/19 9:27 AM