5. To reset the STEL value to zero, press the button.
NOTE:
To cancel the procedure and exit at any time, select
by pressing the left button.
6. Once the value is successfully reset, select by pressing
the right button to return to the Gas options screen.
6.2 BUMP TESTING G6
If configured for your device, bump testing
verifies that your device’s gas sensors and
notification indicators (lights, sound, and
vibration) are functioning correctly. During a
bump test, you apply a known concentration
and amount of gas to confirm the sensor will
trigger a notification due to the gas exposure.
Your bump test schedule should be informed
by your company’s safety policy and is
configurable by your Blackline Live
administrator.
G6 automatically communicates bump test data
to Blackline Live at the device’s next scheduled
synchronization and will remind you when a
bump test is overdue.
For more information on G6 bump testing notifications, refer to Operational Notifications.
To manually bump test G6, you will need a calibration cap and tube. Alternatively, you can
bump test using G6 Dock.