User's Manual

Table Of Contents
©2022 Blackline Safety Corp G6 Technical User Manual | 34
Uncontrolled when printed Rev 01
For more information on bump testing with G6 Dock, see the G6 Dock Technical User Manual on
the Blackline Support Center website.
WARNING: Only perform bump tests in a known clean environment.
To manually bump test G6 using a calibration cap and tube:
1. Attach the tube and calibration cap to G6, ensuring the
tube is securely attached to the fixed flow regulator and
calibration cap.
IMPORTANT:
Do not turn on the gas cylinder until G6
indicates you should do so.
2. From the Home screen, select
by pressing the center
button.
3. Using the navigation buttons, scroll through the Global
Navigation menu and select
Compliance
by pressing the
center button (
).
The Compliance screen opens, displaying information
about when your device’s next bump test and calibration
are due.
4. To access the available
Compliance
procedures, select
by pressing the right button.
The available Compliance procedures display.
5. Using the navigation buttons, scroll to and select
Bump
test
by pressing the center button ( ).
The Bump test screen opens.