G6 stays in the bump test maintenance state while any
residual gas clears.
Once complete, the Compliance screen opens, displaying
information about when your device’s next bump test and
calibration are due
.
IMPORTANT:
If you cancel the bump test or if the bump test
fails, you must turn off the gas and allow any excess gas to
clear prior to disconnecting your device.
If the bump test fails, the device will update to a bump test
overdue status.
If the bump test is cancelled, the device will remain in the
status it was prior to the cancelled bump test.
IMPORTANT:
If G6 is experiencing a sensor error, you will not
be able to perform a bump test until the sensor error is
resolved.
If you see a bump test fail message on your LCD screen, try the bump test again. If the error
persists, please contact our Customer Care team.