37 | G6 Technical User Manual ©2022 Blackline Safety Corp
Rev 01 Uncontrolled when printed
6.3 CALIBRATING G6
If configured for your device, you can manually
calibrate your gas sensor by applying a known
concentration of gas for a set amount of time.
This procedure ensures the gas sensor can
accurately detect gas levels throughout its
operating life. The calibration schedule should
be based on your company’s safety policy.
For more information on G6 calibration
notifications, refer to Operational Notifications.
To manually calibrate G6, you will need a
calibration cap and tube. Alternatively, you can
calibrate using G6 Dock.
For more information on calibrating your device with G6 Dock, see the G6 Dock Technical User
Manual on the Blackline Support Center website.
To manually calibrate G6 using a calibration cap and tube:
1. Attach the tube and calibration cap to G6, ensuring the
tube is securely attached to the fixed flow regulator and
calibration cap.
IMPORTANT:
Do not turn open the gas cylinder until G6
instructs you to.
2. From the Home screen, select by pressing the center
button.
3. Using the navigation buttons, scroll through the Global
Navigation menu and select
Compliance
by pressing the
center button (
).
The Compliance screen opens, displaying information
about when your device’s next bump test and calibration
are due.