©2022 Blackline Safety Corp G6 Technical User Manual | 38
Uncontrolled when printed Rev 01
4. To access the available Compliance procedures, select
by pressing the right button.
The available
Compliance procedures display.
5. Using the navigation buttons, scroll through the procedure
options and select
Calibration
by pressing the center
button (
).
The Calibration screen opens.
NOTE:
To cancel the calibration and exit the workflow, at
any time, select by pressing the left button.
6. To start the calibration workflow, select by pressing the
right button.
G6 performs a hardware self test of your device’s audio,
visual, and vibration capabilities.
G6 zeros the sensor before calibration starts.