User's Manual

Table Of Contents
©2022 Blackline Safety Corp G6 Technical User Manual | 40
Uncontrolled when printed Rev 01
IMPORTANT:
If you cancel the calibration or if the calibration
fails, you must turn off the gas and allow any excess gas to
clear prior to disconnecting your device.
If the calibration fails, the device will update to a calibration
overdue status.
If the calibration is cancelled, the device will remain in the
status it was prior to the cancelled calibration.
IMPORTANT:
If G6 is experiencing a sensor error, you will not
be able to perform a calibration until the sensor error is
resolved.
6.4 ZEROING G6
If configured for your device, you can manually zero your sensors to reset the baseline if G6 is
not reading zero and you know you are in an atmosphere with no gas.
IMPORTANT: If G6’s baseline appears to have shifted, it is best to calibrate your sensors. If you
are unable to perform a calibration, zero your device.
NOTE: The baseline reading for oxygen (O
2
) is 20.9.
To manually zero G6:
1. From the Home screen, select by pressing the center
button.
2. Using the navigation buttons, scroll through the Global
Navigation menu and select
Compliance
by pressing the
center button (
).