User's Manual
8 || G7 EXO TECHNICAL USER MANUAL
2.4 SETUP WIZARD
The setup wizard is an optional test that tells you if EXO will be able
to operate correctly and fully. It determines if EXO can determine its
location, if it can connect to Blackline Live and if it is vertical.
EXO will continue to monitor for gas in the area during the setup wizard
test. Yellow warning alarms, a red latch pull or dangerous gas levels will
override the setup wizard.
To run from start-up
1. Power on EXO
2. EXO will go through the start-up sequence
3. At the end of the start-up sequence a prompt to start the setup
wizard will display on the screen
4. If no selection is made after 15 seconds the prompt screen will time
out and EXO will automatically go to the main gas detection screen
without running the setup wizard
5. If you choose to run the setup wizard, EXO will execute the process
automatically. This process should only take a few minutes.
To run from main menu
1. Press the OK button to enter the main menu
2. Select Setup wizard
3. EXO will execute the process automatically. This process should only
take a few minutes.
If setup wizard is successful,
1. EXO will let you know with a success sound and the screen will
display “ready for use”
2. Select Exit to go to the main gas detection screen
If setup wizard is unsuccessful,
1. EXO will let you know with a failure sound and the screen will
display, “Insucient”
2. The reasons for the failure will be listed on the screen with a red X.
There are three possible reasons for failure:
• EXO could not determine its location,
• EXO could not connect to the Blackline Safety Cloud, or
• EXO was not vertical.
3. Address each item on the list then select Retry to run the setup
wizard again.
4. Select Exit to skip the setup wizard and go to the main gas
detection screen
2.5 CONNECTIVITY LIGHT
Are you connected to the Blackline Safety Cloud?
EXO lets you know its connection status through the green
connectivity light.
Blinking/Solid
Blinking green light
A blinking connectivity light indicates EXO is trying to
connect to the Blackline Safety Cloud. EXO will continue to
monitor the area although monitoring personnel cannot
receive alerts while this light is blinking. Data collected by
EXO while the green connectivity light is blinking will be
sent to Blackline Live once EXO connects to the Blackline
Safety Cloud. This includes red alerts, yellow alarms, location,
messages, etc.
Solid green light
A solid connectivity light indicates EXO is connected to the
Blackline Safety Cloud and all data collected by EXO is actively
being transmitted to Blackline Live. Monitoring personnel will
receive and respond to alerts while this light is solid.
Connection lost alarm
By default, a yellow alarm will be triggered 10 minutes after
EXO loses connection to the Blackline Safety Cloud. If EXO
reconnects to the Blackline Safety Cloud within this time
limit, no alarm will be triggered. This amount of time is
congurable on Blackline Live.
See sections 1.5 and 3.2 for more information on connectivity.