User's Manual

12 || G7 EXO TECHNICAL USER MANUAL
To charge the battery
1. Plug the quick charger into an outlet
2. Lift the rubber ap at the top of the battery pack to expose the
charging port
3. Plug the quick charger into the charging port
4. Charging may take up to 10 hours
NOTE: The battery pack is fully charged when the red light on the
charger turns green
5. When fully charged, remove the quick charger from the charging port
6. Replace the rubber ap to cover the charging port
To insert the battery
1. Ensure EXO is in a safe environment with a clean atmosphere
2. Hold the battery at a 45 degree angle with the bottom pointing
towards EXO
3. Place the battery bottom rst into EXO’s battery slot
4. Push the top of the battery towards EXO until it sits ush
5. Tighten the two screws at the top of the battery
NOTE: The screws should be snug, but not over-tight.
If you see an O2 stabilizing message on G7 EXO’s screen, wait up to
20 minutes for the O2 sensor to stabilize. EXO will NOT monitor during
this process. If this message persists, contact your organizations
safety professional.
TRICKLE CHARGER
You may choose to buy an EXO Trickle Charger Kit from Blackline Safety.
This kit connects EXO directly to a power source through the power
port eliminating the need to power down and remove EXO from the
eld to charge the battery pack. See section 10.1 for more information
about the power port.
SAFETY WARNING: An attached trickle charger is ONLY intrinsically safe
when connected as described in the electrical diagrams in section 10.1.
If you see a O2 stabilizing message on G7 EXO’s screen, wait
up to 20 minutes for the O2 sensor to stabilize. EXO will NOT
monitor during this process. If the message persists, contact your
organizations safety professional.
To power o EXO,
Press and hold the power button for three seconds.
NOTE: If the maintenance code is enabled, you will be required to input
the correct code to unlock EXO before powering o. See section 3.5 for
more information about the maintenance code.
You will know when EXO has nished power o when it completes the
following stages.
EXO will show a three second countdown as you hold the power
button. Every second will have a corresponding beep and ash.
At the end of the three seconds EXO will sound a chime signaling it
is powering o
The screen will go into EXOs shutdown sequence
Once all the lights and sounds have stopped, EXO will be powered
o and disconnected from the Blackline Safety Cloud
NOTE:
Make sure all red alerts have been resolved before powering o.
Do NOT power o EXO if the blue LiveResponce light is on. This may
mean waiting for monitoring personnel to contact you through EXO.
See section 4.2 for more information on red alerts.
Power button
To power on EXO,
Press and hold the power button for
two seconds. You will know when
EXO has nished its startup when it
completes the following stages.
At the end of two seconds EXO
will sound a chime, signaling it is
powering on
The top lights will ash
EXO will go through its startup
sequence and the screen will
display the active features on EXO
The green connectivity light will
stop ashing and become solid
when EXO is connected to the
Blackline Safety Cloud
3.4 POWER ON/OFF