7.2 CALIBRATION
What is calibration?
Calibration ensures EXO can accurately detect gas. This procedure
adjusts sensor parameters while the sensors are exposed to a known
concentration of gas for a set amount of time. Calibrations will need
to be done periodically throughout a sensors operating life and the
data collected from each procedure will be sent to the Blackline Safety
Cloud automatically.
Calibration schedule
The calibration schedule can be congured to match your company’s
safety policy. These changes are made in the conguration prole on
Blackline Live. All the sensors on a cartridge will have the same calibration
schedule, but if you choose to calibrate sensors individually, they can
become due independent of the other sensors. Blackline recommends
calibrating all the sensors on a cartridge in a single calibration process. On
startup EXO will display when the sensors’ next calibrations are due. By
default, an overdue calibration will cause a reminder message to display in
the banner on the gas status screen. This overdue response is congurable
on Blackline Live.
NOTE: Blackline recommends not exceeding 180 days without a calibration.
Gas cylinders
Sensors can be manually calibrated at the same time using one gas
cylinder or individually using multiple gas cylinders. If using multiple
cylinders, the manual calibration process will need to be repeated for
each cylinder. In step 4 of the calibration process below, ensure all the
gas sensors that correspond to the attached cylinders gas mixture
are checked. The gas concentration of the connected gas cylinder
should match the gas concentration listed in EXO’s calibration gas
conguration in Blackline Live.
NOTE: Some cartridges require you to calibrate sensors in a specic
order due to gas sensor cross sensitivity. Contact your safety supervisor
to learn more.
MANUAL CALIBRATION
You will need:
A cylinder(s) containing the appropriate gases
A 0.5 LPM or 1 LPM xed ow regulator attached to the cylinder(s)
Tubing tted with a quick connect coupling insert
To start,
1. Ensure G7 EXO is in clean air
2. Press the OK button to enter the main menu
3. Select Gas options
4. Select Calibration
5. You will see a screen listing the congured calibration settings,
followed by a screen that says, “Continue with calibration?”. Select Yes.
Tubing connected to EXO’s manual calibration inlet