Use Instructions

15
gasoline (except when attached to the Patch and used during a shower or bath).
Do not put the Novii POD into a pocket or bag without adequate protection. Do not
disassemble the Novii POD. Do not crush or pierce the Novii POD. Do not leave
the Novii POD close to a fire or heat source above 30 °C. Do not use the Novii
POD if there are any signs of visible damage. Do not discharge the Novii POD in
any way other than it’s intended use.
Do not use the Novii POD if there is any discoloration, unusual heat, odor or
discharge. Do not put the Novii POD into a microwave or pressurized container.
If liquid leaks from the Novii POD onto your clothes or skin wash well immediately
with fresh water.
If liquid leaks from the Novii POD and comes into contact with your eye, do not rub
your eye, wash well with clean edible oil and see a doctor immediately.
WARNING: Do not charge the Pods on an external wireless charger, only charge via the Novii
Interface
CAUTION: Keep the operating environment free of dust, vibrations, corrosive, or flammable
materials, and extremes of temperature. The Novii Interface and all cable
connectors should be kept clean and free of electrode gel and other substances.
CAUTION: The Novii Interface is rated IPX0. Do not operate the Novii Interface if it is damp or
wet because of condensation or spills. Avoid using the equipment immediately
after moving it from a cold environment to a warm, humid location.
CAUTION: The Novii POD on its own is rated IPX0. The Novii POD is rated IP57 only when
mated with the Novii Patch. Do not submerse the Novii POD in any liquid if not
mated to a patch.
CAUTION: Never use sharp or pointed objects to operate the touch screen display. Do not
exert excessive pressure when operating the touch screen.
CAUTION: The POD gold connection pins need to be kept clean, and should be protected at
all times; only keep your PODs in the Interface charging bays or clipped to a Patch.
Placing it down anywhere else could result in damage to the gold pins.
2.4 Electromagnetic Compatibility (EMC)
2.4.1 Electromagnetic Interferences
The Novii System has been designed to minimize the impact of electromagnetic interference
from other electrical equipment and also to minimize the interference caused to other electrical
equipment by the Novii System. The Novii system has been tested and found to comply with the
Medical Electrical Equipment - General Requirements for Safety-Collateral Standard:
Electromagnetic Compatibility, EN60601-1-2:2007, and FFC CRF47 Parts 15.107 & 15.109,
Class A limits. However because of proliferation of radio-frequency transmitting equipment and