5.3.2 DISPLAY MHR ON INTERFACE
Touching this item ‘bar’ will Enable or Disable the MHR display on the Novii Interface.
Selecting to display the MHR on the Novii Interface will automatically turn on the “MHR/FHR
coincidence Alert”. The default is not to display the MHR on the Novii Interface. As well as a
visual alert there is also an audio alert and this will be enabled if the AUDIO ALERTS are turned
ON, see Section 5.3.3 below.
5.3.3 AUDIO ALERTS
The factory default is AUDIO ALERTS DISABLED and can only be changed in the SETUP. By
touching the AUDIO ALERTS item ‘bar’ in SETUP the audio alerts can be ENABLED, providing
an audible alert to supplement the visual alert for the following situations:
i. Low POD battery - Audio alert is always enabled
ii. POD not returned to Interface charging bay - Audio alert is always enabled
iii. MHR coincident with FHR (only if the DISPLAY MHR ON INTERFACE has been
Enabled) and Audio Alerts have been enabled
iv. Electrode(s) detached from abdomen. Audio alerts need to be enabled
v. Patch not genuine - Audio alert is always enabled
Once an alert sounds it can be silenced by touching the SOUND icon which will be flashing or
by following on screen instructions. If the alert condition continues the alert will repeat according
to the schedule below:
Alert Condition
Initial Alert Condition
Once acknowledged Audio Alert
will repeat if the condition does
not resolve after
Low battery
Up to 60 minutes battery life left
15 minutes
MHR coincident with FHR
MHR is within ±10 bpm of FHR for
60 seconds
60 minutes
POD left in Patch and Novii
Patch electrode/skin
preparation check is not
passed or bypassed
After 10 min
Will not be repeated once alert has
been cancelled
POD not returned after
removed from Patch
2 minutes after end of 2 minute
count down
Will not repeat after POD is docked
or alarm condition is acknowledged
on display screen
POD not attached to Patch
After 2 minute count down has
finished
Will not repeat after POD is docked
or alarm condition is acknowledged
on display screen
Electrode(s) detached from
abdomen
When electrode(s) detached
Will not repeat after audio alert has
been silenced