5.8 TEST function
To confirm that the Novii Interface, Maternal/Fetal Monitor, GE Y adaptor and cables work
correctly, touch the TEST icon from the Start screen. A signal will be sent to the Maternal/Fetal
Monitor to check correct functionality. Monica recommends that whenever the user requires
evidence to demonstrate the correct operation of the Interface and Maternal/Fetal Monitor e.g.
after installation, or to confirm that there are no breaks in the cables or a fault has developed;
the TEST icon on the Start screen should be used. The GE Y adaptor should always be moved,
shaken, to ensure there are no intermittent problems.
After touching the TEST icon the user will be asked to zero the TOCO on the Maternal/Fetal
Monitor see below, and confirm using the forward arrow key.
Once confirmed, a test FHR, MHR and UA signal will be sent from the Novii Interface to the
Maternal/Fetal Monitor. The FHR, UA and MHR (if connected) values displayed on the
Maternal/Fetal Monitor digital display should match the FHR, MHR and UA numbers displayed
on the Novii Interface display: