Use Instructions

44
c. Position POD clip so that the top edge of electrode #2 is 5”/10-12cm below the
fundus.
The electrode on the flexible cable, electrode #5, Fig. 2, should be placed on top of
pannus approximating to the point 2.4”/6cm horizontally from the symphysis pubis looking
vertically down. This is difficult to estimate and if the FHR signal is poor, reposition this
electrode lower down on abdomen to maximize FHR signal and consider
placing under the pannus just below turn ensuring the electrode is
not folded.
6.5.2 Instruction 1: Zero UA on Maternal/Fetal Monitor
Press the UA zero reference button on theFetal Monitor.
6.5.3 Instruction 3: Select a Charged POD
1. Remove any Novii POD from one of the Novii Interface charging bays as long as the POD battery
status icon is GREEN. Once it is removed the blue lights on the front of the Novii POD will flash
alternately, to indicate that the POD is now ‘active’ and paired to the Novii Interface.
2. The Interface display will change to a countdown as shown below. The Novii POD must now
be clipped to the Patch, within 2 minutes.
3. The battery charging icon on the Interface will be replaced by a ‘busy’ icon ( 1, 2, 3 white dots),
indicating that the POD is preparing to commence monitoring. The busy icon will remain until
the MHR is detected when it will stop.
4. The POD is attached to the Patch with the Monica symbol facing up. Magnets in both the
Patch clip and POD ensure correct placement and, no force is required.