Use Instructions

46
10. If you need to end the setup, or monitoring session, remove the POD from the Patch and return
it to the charging bay on the Novii Interface that it came from.
11. The monitoring screen helps the user achieve the best signal quality, control the monitoring
mode, view status alerts and if enabled display the MHR. The format of the display is shown
above.
12. Once the POD identifies and extracts the MHR both blue lights will flash together every 2
seconds.
13. FHR, MHR and UA monitoring should commence within one minute, once the monitoring screen
above is displayed.
14. FHR, MHR and UA data is collected wirelessly from the Novii POD and sent to the Novii Interface
and then on to the Maternal/Fetal Monitor via the Novii Interface cables. The Maternal/Fetal
Monitor acts as if a FECG scalp cable, MECG cable and a TOCO/IUPC transducer cable are
connected and will display and print the FHR, MHR and UA. The user can swap from one or
more of the Monica monitoring modalities to another method e.g. Monica UA, simply by removing
the Novii UA interface plug and replacing with the TOCO UA transducer plug.
WARNING: This is not supported or recommended because the US FHR and TOCO/IUPC UA
are not synchronized with the Novii FHR, MHR and UA, see Section 6.12. It is
recommended that a note is made on the trace or patient notes. Please refer to
the Maternal/Fetal Monitor manufactures instructions for more details on the
display and printing options available.
15. If ‘Display MHR on Interface’ option is enabled in the settings even if the mECG cable supplied
with the Novii Interface is connected to the Maternal/Fetal Monitor, a digital display of MHR will
be shown on the Novii Interface.