Maternal/Fetal Monitoring Troubleshooting Table continued
ID
Symptom
Description
Possible Causes
Actions and Solutions
S8
FHR/UA or MHR
data not being
displayed by
Maternal/Fetal
Monitor
POD problems
Make sure the monitoring screen on the Interface is
being displayed and the signal quality, x3 green
squares, and the Battery Icon below the signal quality
is green. If not confirm the status of the two blue lights
on the POD connected to the Patch. If no blue lights
are visible, the POD has switched off. Remove from
Patch and place in a charging bay. Is it recognised by
the Interface (battery status above POD will appear)?
If not, wait 20minutes, if POD is still not recognised
replace POD. If the battery status was orange wait for
battery to charge. When battery status is green take
POD and place in Patch Clip. If electrode check is
good, but the POD switches off again, replace POD.
Was the ‘Bypass’ button on the Interface electrode
check screen used. If yes, remove POD from Patch
and place in charging bay. Wait a few moments for the
POD to be recognised (battery status above POD will
appear). If this does not happen replace the POD. If it
is recognised remove POD and place in Patch. Ensure
that all electrodes pass electrode check.
S9
FHR quality on the
Maternal/Fetal
Monitor trace is
poor in some
patients
Unfortunately this can happen in some
patients especially during stage 2.
Unless it is persistent and occurs on most/all patients it
is not a fault. The user should follow the Alert/Help
message on the Novii interface. If the FHR is
intermittent, the FHR can be confirmed with the
Doppler Transducer connected to US 2, but if the
problem persists then we would advise removing the
Novii POD/Patch and swapping to another monitoring
modality.