User's Manual

Environmental requirements
Vscan Air
User Manual
3-7
Direction GP092020-1EN
Rev 08
Environmental requirements
Environmental requirements for Vscan Air CL
Transient operating conditions
NOTE: Permissible transient environmental operating conditions:
Temperature range of -20 to +50 degrees
Device will function for a minimum of 20 minutes when
placed in an environment with temperatures ranging from
-20°C to 50°C after storage at room temperature (20 ± 2°C)
Following storage at temperatures ranging from -20°C to
+50°C, the device will within 10 minutes after being returned
to room temperature (20 ± 2°C), function for a minimum of
20 minutes
NOTE: Avoid exposing the unit to saline moisture. In case of exposure
to saline moisture, clean the unit as described in ‘Cleaning and
disinfection’ on page 6-4.
Image on the display device hosting the Vscan Air app is
dependent on ambient light; avoid direct sunlight or reflections
from other light sources on the display when scanning and
reviewing images. The display viewing angle should be as small
as possible.
If you are having difficulty seeing the image due to the lighting
conditions try to change brightness of the display device or
change your position/location of use.
Table 3-1: Environmental requirements
Description Operational Non operational Storage and transport
Temperature 0°C to + 35°C - 40°C to + 70°C - 40°C to + 70°C
Humidity 10–90% 10–95% 10–95%
Air pressure 54 kPa to 106 kPa 50 kPa to 106 kPa 50 kPa to 106 kPa