Vscan Extend Maintenance
6-4 Vscan ExtendUser Manual
5721203-100
Rev. D
Cleaning and disinfection
Cleaning the device
Make sure the device is disconnected from the AC/DC adapter
before cleaning.
1. Power off the Vscan Extend.
2. Moisten a soft, non-abrasive cloth with a mild, general
purpose, non-abrasive soap and water solution.
3. Wipe the Vscan Extend.
4. Wipe dry with a soft towel.
Reprocessing recommendation (Frequency)
After Each Use
Inspect the transducer (see ‘Inspecting the Vscan Extend’
on page 6-3)
Clean the transducer (see ‘Disinfecting the probe’ on
page 6-9).
If required, clean the device and the display (see
‘Disinfecting the device’ on page 6-7).
If required, disinfect the transducer (see ‘Disinfecting the
probe’ on page 6-9).
Ensure that the main unit and the transducer are properly
cleaned after each use and before storage in the case.
CAUTION
Do not spray the soap and water solution directly onto the
Vscan Extend.
WARNING
If any defects or damages are found on the probe or its cable,
DO NOT use the Vscan Extend. Contact GE service.