User's Manual

Vscan Extend Maintenance
6-20
Vscan Extend
User Manual
5721203-100
Rev. 13
Association request failed (or never connected). Request to connect to DICOM Server failed.
Contact Hospital IT team.
Network timeout error Check network connection.
Contact Hospital IT team.
Peer aborted association Repeat the DICOM transaction. If it consistently
fails, contact GE Service.
No network connection Check network connection.
Contact Hospital IT team.
Refused - Server out of resources The DICOM endpoint server is out of resources to
accept exported DICOM file.
Contact Hospital IT team.
Identifier does not match SOP class DICOM file's data set does not match with the SOP
Class.
Contact GE Service.
Failure - Processing failure Interoperability issue.
Contact GE Service.
Failure - Invalid object instance Contact GE Service.
Internal application error Contact GE Service.
Table 6-3: DICOM messages
User message Remarks