User's Manual

Introduction
Vscan Extend
User Manual
8-5
5721203-100
Rev. 13
Audit logging content
The following events are captured by the audit logging of the
Vscan Extend:
1. Device Start/Shut-down time
2. Incorrect password attempts
3. Addition and Modifications to system configuration,
including:
a. DICOM connection
b. Windows Share connection
c. Application software installation/uninstallation/software
upgrade
d. Registration of partnering apps
4. Events related to Patient Data, including details of:
e. Data access Date/Time including type of action
(addition, deletion, modification, reviewed) excluding
the PC connect scenario
f. Details of applications used (Bladder app, Lung
protocol, etc.)
5. Data Export
a. Data export Date/Time including end point and Wi-Fi
SSID information
b. Data Backup/Restore detail
6. Invalid Device PIN attempt
7. Kiosk admin mode login success and failure
Management of Audit logs
Back up of audit logs
Back up of audit logs can be done by exporting audit logs to an
external device. The export happens through USB export.
Remember that audit logs and exported audit log files do contain
Personal Information (PI) and must be handled according to
applicable regulations and guidelines for handling of PI/PHI.
Note that exported audit log files are stored unencrypted.