Electromagnetic emissions
Electromagnetic immunity
Table 2-1: Electromagnetic emissions
Guidance and manufacturers declaration – electromagnetic emissions.
The Vscan Extend is intended for use in the electromagnetic environment below. The
customer or the user of the Vscan Extend should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emission
EN55011
Group 1 The Vscan Extend uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emission
EN55011
Class B The Vscan Extend is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emission
EN/IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
EN/IEC 61000-3-3
Complies
Table 2-2: Electromagnetic immunity (Part 1)
Guidance and manufacturer’s declaration – electromagnetic immunity.
The Vscan Extend is intended for use in the electromagnetic environment below. The
customer or the user of the Vscan Extend should assure that it is used in such an
environment.
Immunity test
EN/IEC 60601
test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
EN/IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst
EN/IEC 61000-4-4
±2 kV for power-supply
lines
±1 kV for input/output
lines
±2 kV for
power-supply lines
±1 kV for input/
output lines
Mains power quality should
be that of a typical
commercial or hospital