Acclimatization time
Allow the unit to acclimate for approximately 10 minutes, if
stored at temperatures ranging from -20°C to +50°C.
Allow the unit to acclimate for approximately 30 minutes, if
stored at temperatures outside the transient environmental
operating conditions -20°C to +50°C.
Other environment: Aircraft Ambulance/Road Ambulance
The Vscan Extend can be used in an emergency medical
services environment, including road and air ambulances*,** .
• The device must not be mounted or fixed inside any
ambulance.
• The USB and Service port doors must be closed while
operating the device.
• The device must be put into flight mode/Wi-Fi off.
• The device must be operated according to the
environmental requirements. See ‘Environmental
requirements for the device’ on page 3-4 for more
information.
• Do NOT connect the device to any outlets available in a
road or air ambulance.
• The device MUST NOT be operated above 4,000 meters.
* The device is compliant to IEC60601-1-12 and EN13718 as
stated in Table i-1. Additional regulations might apply.
** The device is compliant to IEC60601-1-12 and EN1789 as
stated in Table i-1. Additional regulations might apply.
Other environment: Home healthcare
The USB and Service port doors must be closed while operating
the device.