Environmental requirements
Vscan Extend
User Manual
3-5
5721203-100
Rev. 13
Acclimatization time
Allow the unit to acclimate for approximately 10 minutes, if
stored at temperatures ranging from -20°C to +50°C.
Allow the unit to acclimate for approximately 30 minutes, if
stored at temperatures outside the transient environmental
operating conditions -20°C to +50°C.
Other environment: Aircraft Ambulance/Road Ambulance
The Vscan Extend can be used in an emergency medical
services environment, including road and air ambulances*,** .
The device must not be mounted or fixed inside any
ambulance.
The USB and Service port doors must be closed while
operating the device.
The device must be put into flight mode/Wi-Fi off.
The device must be operated according to the
environmental requirements. See ‘Environmental
requirements for the device’ on page 3-4 for more
information.
Do NOT connect the device to any outlets available in a
road or air ambulance.
The device MUST NOT be operated above 4,000 meters.
* The device is compliant to IEC60601-1-12 and EN13718 as
stated in Table i-1. Additional regulations might apply.
** The device is compliant to IEC60601-1-12 and EN1789 as
stated in Table i-1. Additional regulations might apply.
Other environment: Home healthcare
The USB and Service port doors must be closed while operating
the device.