User's Manual

Preparing Vscan Extend for Use
3-24
Vscan Extend
User Manual
5721203-100
Rev. 13
Battery specifications
In order to get maximum charging capacity with your Vscan
Extend battery, you should initially allow the battery to be fully
charged and then fully discharged at least three times. Perform
normal operation during these cycles. Once the initial charging/
discharging cycles are performed, the following is applicable
without reducing the lifetime of the battery:
It is not necessary to completely discharge the battery
before re-charging.
It is possible to stop charging the battery before it is fully
charged, but the battery will then be discharged more
rapidly.
It is possible to charge the battery several times each day, if
needed.
When storing the Vscan Extend for a period longer than three
months, remove the battery from the Vscan Extend main unit.
Battery should be charged every 3 months for long term storage.
NOTE: Make sure to remove the battery from the device before
shipping from one place to another.
To minimize battery performance degradation, avoid prolonged
storage of the battery outside the Vscan Extend operational
temperature range (see ‘Environmental requirements for the
device’ on page 3-4).
Table 3-9: Battery specification
Item Specification
Charging time at >=90% About 75 minutes
Capacity About 1 hour while continuously scanning
Lifetime At least 300 charges