User Manual Part 4

1-10 ApexPro™ 2001989-301A- draft 1
Introduction
WARNING
LOSS OF DATA —Caregivers and telemetry patients should be
made aware of antenna coverage areas. Movement outside of the
coverage area may result in loss of patient monitoring.
WARNING
NETWORK INTEGRITY — The ApexPro server resides on the
Unity network. It is possible that inadvertent or malicious network
activity could adversely affect patient monitoring. The integrity of
the Unity network is the responsibility of the hospital.
WARNING
OPERATOR — Medical technical equipment such as this monitor/
monitoring system must only be used by persons who have received
adequate training in the use of such equipment and who are capable
of applying it properly.
WARNING
POWER SUPPLY — All devices must be connected to a properly
installed power outlet with protective earth contacts only. If the
installation does not provide for a protective earth conductor,
disconnect the device from the power line and operate it on battery
power, if possible.
All devices of a system must be connected to the same power supply
circuit. Devices which are not connected to the same circuit must be
electrically isolated when operated (electrically isolated EIA232
interface).
WARNING
PROTECTED LEADWIRES — Only use protected leadwires and
patient cables with this device. The use of unprotected leadwires and
patient cables creates the potential for making an electrical
connection to ground or to a high voltage power source which can
cause serious injury or death to the patient.
322C
Unprotected Leadwire
Protected Leadwire