User Manual Part 5

2-4 ApexPro™ 2001989-301A- draft 1
Equipment overview
Optional components
Apex oximeter
CAUTION
Do not use the Apex oximeter on neonatal patients. It is not designed
for use on neonates.
An Apex oximeter can be connected to a transmitter in order to monitor the patient’s
pulse oximetry data and send the SpO2 data for display at the CIC Pro center. Only
digital data is available; no waveforms are generated or transmitted. Digital data is
stored in Graphic Trends and Vital Signs. For setup information, refer to Apex
oximeter on page 3-7.
Xpod oximeter
An Xpod oximeter can be connected to a transmitter in order to measure arterial
oxygen saturation (SpO2), peripheral pulse rate (PPR), and perfusion quality and send
the data for display at the CIC Pro center. For setup information, refer to Xpod
oximeter on page 3-7.
Accutracker DX noninvasive blood pressure (NBP) monitor
NOTE
The Accutracker DX noninvasive blood pressure monitor has been modified by
SunTech Medical Instruments to operate with the ApexPro system.
An Accutracker DX noninvasive blood pressure monitor can be connected to a
transmitter in order to measure the systolic and diastolic blood pressures and send for
display at the CIC Pro center. Digital values are stored in Graphic Trends and Vital
Signs. For setup information, refer to Accutracker DX on page 3-10.
Dinamap PRO monitors
NOTE
The Dinamap PRO monitors’ alarm limits are not configurable at the CIC Pro
center, but they can be silenced at the CIC Pro center. However, alarms that are
silenced at the CIC Pro center will not be silenced at the monitor. Refer to the
Dinamap Pro 100–400 Operation Manual for detailed information.
A Dinamap PRO 100, 200, 300, and 400 monitor can be connected to a transmitter in
order to monitor SpO2, NBP, and temperature and send the data for display at the CIC
Pro center. For setup information, refer to Dinamap PRO monitors on page 3-9.
Draft