User Manual Part 6

6-6 ApexPro™ 2001989-301A- draft 1
Managing patients
883B, 884B
Verify status
After the transmitter leadwires have been properly attached to the patient's electrodes,
verify lead quality, electrode status and transmitter function.
For more information, refer to Maintaining quality ECG signal on page 6-6.
For more information, refer to Transmitters on page 2-2.
Maintaining quality ECG signal
Electrodes are disposable and applied only once. Attempts to replace a loose
electrode guarantees excessive motion artifact and results in false alarms. Regardless
of patient type, electrodes should be replaced at least every 48 hours to maintain
quality signals during long-term monitoring. Over the course of 48 hours, the
electrode gel will start to dry out and the adhesive will age. This may irritate the
patient’s skin.
The transmitter must be carried securely on the patient. Stabilize the electrode and
leadwire with a leadwire stress loop near the electrode. Tape the stress loop to the
patient. A secured stress loop prevents the leadwire rotation about the electrode snap,
leadwire tugging at the electrode and ECG artifact.
306A
If the transmitter is allowed to dangle or jostle as the patient moves, then the weight
of the device will tug on the electrodes, causing degradation of the signal. Even if the
electrodes are in good contact, excessive tugging on the electrodes will change the
physical structure of the gel, thereby generating motion artifact signals. Transmitters
Lead II
Lead II
Lead I
Lead I
Neonatal AHA Electrode Placement
Neonatal IEC Electrode Placement
Draft