User Manual Part 6

System setup
2001989-301A- draft 1 ApexPro™ 4-15
Full disclosure defaults
This option sets the full disclosure settings. In user mode, only the full disclosure
Report and Strip settings can be configured. You must be in the Service mode to set
the other full disclosure settings at the CIC Pro center.
To view the full disclosure default settings, click CIC Setup > Full Disclosure
Defaults.
503A
Option Function
Report
Duration Designate how much data is included in the report. The
maximum report duration is 72 hours, depending upon
licensing.
To set the report duration, place the cursor on the scroll bar
below the Report Duration display field. Move the scroll bar
to the left for shorter duration or to the right for longer
duration.
Hole Location Provide space for binding printed reports. Choices are:
none, top, bottom, left, and right.
Include Set print characteristics. You may set any or none of these
options. Choices are: Graybar, Arrhythmia Annotations,
and Heart Rate.
Line Time Designate how much data shows on an individual report
line. Choices are: 15sec, 30sec, and 1min.
Strip
Draft