User Manual Part 7

Viewing real-time patient data
2001989-301A- draft 1 ApexPro™ 7-17
Single-Lead ECG telemetry data
NOTE
When acquiring Single-Lead ECG data using a 5- or 6-leadwire set, it is not
necessary to connect the V leads or the right leg lead to the transmitter or to the
patient.
The following constraints apply when using Single-Lead ECG telemetry data.
Arrhythmia
WARNING
INCORRECT ALGORITHMS, ARRHYTHMIA PROCESSING
AND CALCULATIONS BASED ON PATIENT AGE — After
manually updating or automatically retrieving patient demographic
information from a network database, always confirm that the
entered patient’s date of birth matches the patient’s actual date of
birth. Otherwise the appropriate age-related algorithms, arrhythmia
detection, and calculations will not be applied.
Function Single-Lead Constraints
change the displayed
lead
The factory default Display Lead is lead II.
Contact your local service representative to change the
default displayed lead.
Display Lead appears to be selectable at the CIC Pro
center. However, your selection is temporary and will
revert back to the transmitter’s default displayed lead.
NOTE
When the clinical situation dictates monitoring a lead
other than the default lead, you can move the leads
and/or electrodes to view a different lead. Be aware
that the label on the display and on the printout will
show the default lead label.
Lead Analysis Multi-Lead analysis may appear to be selectable at the
CIC Pro center. However, your selection is temporary and
will revert back to the Single-Lead analysis mode.
select a V lead V leads may appear to be selectable at the CIC Pro
center. However, your selection does not change the
transmitter’s default displayed lead.
select displayed leads
from a single viewer
Leads other than the default displayed lead may appear to
be selectable at the CIC Pro center. However, your
selection is temporary and will revert back to the
transmitter’s default displayed lead.
select graph waveforms Leads other than the default displayed lead may appear to
be selectable at the CIC Pro center. However, you must
select the transmitter’s default displayed lead to obtain a
graph of the waveform.
Draft