User Manual Part 7

Viewing real-time patient data
2001989-301A- draft 1 ApexPro™ 7-29
5. After making your selections, complete one of the following tasks:
Click a different Monitor Setup option to apply your changes without
closing the Monitor Setup window.
Click the (close button) on the top right side of the window to apply
your changes and close the Monitor Setup window.
NBP monitoring with telemetry
NBP monitoring via telemetry is done with an Accutracker DX or Dinamap Pro blood
pressure monitor connected to the transmitter. The blood pressure cuff is connected to
the blood pressure monitor, which measures and displays systolic and diastolic blood
pressures using the auscultatory method. When the blood pressure monitor is
connected to an transmitter, digital values are also displayed at the CIC Pro center.
WARNING
The following conditions may affect the accuracy of noninvasive
blood pressure readings: seizures, tremors, extreme hypotension or
hypertension, arrhythmias, or extremely high or low heart rate.
Patient preparation
Blood pressure cuff selection and application are important. Inappropriate selection or
improper application of the cuff will result in erroneous measurements.
Most people use their non-dominant arm for acquiring ambulatory noninvasive blood
pressure readings.
Follow these steps to prepare the patient for NBP monitoring:
Non-invasive blood pressure control settings
Option Function
Auto Turn Off automatic NBP measurements.
When turned On at the monitor, NBP measurements are
acquired automatically at regular intervals.
NOTE
This option does not apply to telemetry beds and
cannot be turned on from the CIC Pro center.
Cuff Size Set the inflation pressure used during the first NBP
measurement and for calculating the NBP pressure:
NOTE
For more information, refer to the appropriate
bedside monitor’s operator’s manual.
Clear Message button Clear the display of inflation messages and current NBP
readings.